![background image](http://html1.mh-extra.com/html/omega/hhm14/hhm14_user-manual_3993732006.webp)
4
DC CURRENT
Ranges: 200µA,2mA,20mA,200mA,10A
Accuracy: ±(0.5%rdg + 10dgts) on 200µA to 200mA ranges
±(2.0%rdg + 10dgts) on 10A range
Input protection: 0.5A / 250V fast blow fuse
10A / 600V fast blow ceramic fuse
AC CURRENT (True RMS 50Hz - 500Hz)
Ranges: 200µA,2mA,20mA,200mA,10A
Accuracy: ±(1.0%rdg + 10dgts) on 200µA to 200mA ranges
±(2.5%rdg + 10dgts) on 10A range
Input protection: 0.5A / 250V fast blow fuse
10A / 600V fast blow ceramic fuse
RESISTANCE
Ranges: 20
W
,200
W
,2K
W
,20K
W
,200K
W
,2000K
W
,20M
W
Accuracy: (Fuse resistance is not included)
±(1.0%rdg + 200dgts) on 20
W
range
±(0.2%rdg + 20dgts) on 200
W
range
±(0.2%rdg + 2dgts) on 2K
W
to 2000K
W
ranges
±(2.0%rdg + 3dgts) on 20M
W
range
Open circuit volts: < 3.5V on all ranges
Overload protection: 250VDC or AC rms on 20
W
range
500VDC or AC rms on other ranges