SAFETY  INFORMATION

  The following safety information must be observed to ensure maximum
personal safety during the operation of this meter:

1. Do not use the meter if the meter or test leads look damaged, or if you

suspect that the meter is not operating properly.

2. This meter is not recommended for high voltage industrial use; for example,

not for measurements of 440 VAC or 600 VAC industrial power mains. The
unit is intended for use with low energy circuits to 1000VDC or 750VAC or
high energy circuit to 250 VAC or DC. Accidental misuse by connection
across a high voltage, high energy power source when the meter is set up
for mA measurement may be very hazardous.

3. Turn off power to the circuit under test before cutting, unsoldering, or

breaking the circuit. Small amounts of current can be dangerous.

4. Use caution when working above 60V dc or 30V ac rms. Such voltages pose

a shock hazard.

5. When Using the probes, keep your fingers behind the finger guards on the

probes.

6. Measuring voltage which exceeds the limits of the multimeter may damage

the meter and expose the operator to a shock hazard. Always recognize the
meter voltage limits as stated on the front of the meter.

7. If the equipment is used in a manner not specified by the manufacturer, the

protection provided the equipment may be impaired.

Summary of Contents for HHM14

Page 1: ...OMEGA HHM14 Digital Multimeter...

Page 2: ...l Free in Germany 0800 82 66342 e mail nl omega com e mail germany omega com Czech Republic United Kingdom ISO 9002 Certified ul Rude armady 1868 733 01 Karvina One Omega Drive Hranice Czech Repubic R...

Page 3: ...across a high voltage high energy power source when the meter is set up for mA measurement may be very hazardous 3 Turn off power to the circuit under test before cutting unsoldering or breaking the...

Page 4: ...ating Environment 0 C to 40 C at 70 relative humidity Storage Temperature 20 C to 60 C 0 to 80 R H with battery removed from meter Accuracy Stated accuracy at 23 C 5 C 75 relative humidity Safety Acco...

Page 5: ...AC VOLTS True RMS Ranges 200mV 2V 20V 200V 750V Resolution 10 V Accuracy Input impedance 10MW Overload protection 500VDC or 350VAC 30 sec on 200mV range 1000VDC or 750VAC rms on other ranges Range Inp...

Page 6: ...10dgts on 200 A to 200mA ranges 2 5 rdg 10dgts on 10A range Input protection 0 5A 250V fast blow fuse 10A 600V fast blow ceramic fuse RESISTANCE Ranges 20W 200W 2KW 20KW 200KW 2000KW 20MW Accuracy Fu...

Page 7: ...40W 20W Overload protection 500VDC or AC rms DIODE TEST Test current 1 0mA 0 6mA Accuracy 2 0 rdg 3dgts Open circuit volts 3 3Vdc typical Overload protection 500VDC or AC rms TRANSISTOR hFE Ranges 0 1...

Page 8: ...itch to the highest range and reduce until a satisfactory reading is obtained 3 Connect the test leads to the device or circuit being measured 4 Fordc a signisdisplayedfornegativepolarity positivepola...

Page 9: ...er from the equipment under test 3 Connect the red test lead to the VW jack and the black test lead to the COM jack 4 Touch the probes to the test points In ohms the value indicated in the display is...

Page 10: ...both directions 7 If the junction is measured in a circuit and a low reading is obtained with both lead connections the junction may be shunted by a resistance of less than 1kW In this case the diode...

Page 11: ...are possible check for a blown overload protec tion fuse There are two fuses F1 for the mA jack and F2 for the 10A jack For access to fuses remove the two screws from the back of the meter and lift o...

Page 12: ...ty or otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipmen...