The HHF1000 series air velocity/temperature meter is 

not explosion proof, nor is it intrensically safe. Do not 

use for flammable or hazardous gases, or in hazard-

ous areas.

The HHF1000 series air velocity/temperature meter 

is intended for use with clean air or Nitrogen ONLY. 

Do not use with other gases, as it will produce an 

un-calibrated and non-linear display measurement. 

In addition, air carrying dust or oil (such as found in 

blow/compressor systems that utilize oil) can lead to 

coating of the sensor and thus inaccurate readings.
The HHF1000 is a bi-directional device, meaning the air 

flow in the forward or reverse direction provides the 

same readings.

Installation

1.  Install the 9V battery in the battery compartment of 

the case. You can remove the battery door to get to 

the compartment.

2. Remove the protective cap from the sensor tip.
3.  Run a length of straight pipe before and after the 

flow sensor probe. The amount of upstream straight 

pipe required depends on the type of obstruction 

which is immediately upstream of the flow sensor. 

Downstream of the flow sensor, in all situations, 

run 5 diameters of straight pipe regardless of the 

downstream obstruction.

4.  Align the sensor probe with the air flow. Make sure 

the air flow is perpendicular to the sensor windows. 

The score line on the sensor tubing is another way of 

aligning the sensor to the flow stream. The score line 

starts from the center of the sensor window and as a 

result it can be aligned properly.

5.  One way of installing the sensor probe into a flow 

stream is to utilize a compression fitting such as 

Omega’s SSLK-14-14 stainless steel compression 

fitting with Teflon ferrule, which allows adjustment 

of the insertion depth of the probe.

6.  Turn on the power switch and start measuring air 

velocity & temperature.

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

START HERE

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

2

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

3

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

4

Using This Quick Start Manual

Use this Quick Start Manual with your HHF1000 

series handheld air velocity/temperature meter for 

quick installation and basic operation. For detailed 

information, refer to the User’s Guide (Manual 

Number M4979).

General Information

The HFF1000 series handheld air velocity/tem-

perature meter measures and displays air velocity 

mass flow and air temperature of clean air flows 

in ducts & pipes, while producing very little pres-

sure drop in the flow stream. The sensor design is 

based on three RTD elements, one measures the 

air temperature and the other two measure the air 

velocity. 
The HHF1000 displays the air velocity in feet per 

minute (FPM), meter per second (m/s), miles per 

hour (MPH), and kilometer per hour (Km/h). The 

air temperature is displayed in °F & °C. The vol-

ume air flow is displayed in Cubic Feet per min-

ute (CFM) and Cubic meter per minute (CMM). 

The sensor probe is 12" long as standard. The 304 

Stainless steel sensor tubing is provided with inch 

marks for ease of insertion depths. The sensor 

probe comes in two different versions as follows:

• Fixed 12 inches long Probe

• Telescopic probe adjustable from 10 to 36 inches. 

The unit powers from a 9V battery or a 9Vdc 

adapter.

Piping Requirement

Main Operation

Changing Display Engineering Units – You can 

change the air velocity Engineering unit display from 

feet per minute (FPM) to meter per second (m/s), miles 

per hour (MPH), and kilometers per hour (Km/h) 

from the keypad (Press SET key). You can change the 

temperature display from °F to °C or vice versa by 

pressing the DOWN key.
Turn On Display Backlight – You can turn the 

LCD backlight on or off by pressing the UP key. The 

backlight will turn off after 10 seconds if the UP key is 

not pressed again.
High & Low Alarm Set Points – You can set and 

enable the high & low alarm set points for the air 

velocity as shown in the keypad flow chart. When in 

the alarm condition, the HAL or LAL icon will flash on 

the LCD.

Display Volume Air Flow – The unit calculates 

volume air flow by multiplying the air velocity by 

the cross sectional area of the pipe or the duct. The 

cross sectional area can be set in the display menu 

as shown in the keypad flow chart in square inches 

units. The volume air flow is shown in cubic feet 

per minute or cubic meter per minute when air 

temperature is displayed in Degrees C.
Auto Power Shutoff – You can turn on the auto 

power shutoff feature (Default is ON) to save 

battery life. If no keys are pressed for 5 minutes, 

the unit will shut down to save battery life. 

Pressing any key will turn the unit back on.

Keypad Functional Flow Chart

The HHF1000 air velocity/temperature meter 

saves the following parameters in the non-

volatile memory, so removing the power will not 

affect these settings: High and Low alarm air 

velocity set points and status (On or Off), Display 

Response time in msec, Cross sectional area in 

Square inches, Atmospheric Pressure in mmHg.

NOTE:

NOTE:

 

Real Time 

Go to Max  

FPM       m/s 

°F       °C or 

Turn On/off 

   

Mode 

 

vice versa 

LCD Light 

   

 

Km/h         MPH

Display Max Vel.   Go to Min  

   ____________ 

    _____________ 

Reset Max, Min,  

& Temp MAX Icon  Mode 

 

 

Velocity &  

  

 

 

 

Temperature

Display Min Vel.   Go to Volume  

   ____________ 

    _____________ 

Same 

& Temp MIN Icon  Air Flow Mode

Display Vol. Flow   Go to High Alarm     ____________ 

    _____________ 

   ____________ 

In CFM or CMM 

Mode Velocity

Display High  

Go to Low  

Enable/Disable 

Increment high 

Decrement high 

Alarm Velocity 

Alarm Mode  

High Alarm, HAL 

Alarm Set Point 

Alarm Set Point 

Set Point 

Velocity 

Icon when enabled

Display Low  

Go to Display 

Enable/Disable 

Increment Low 

Decrement Low 

Alarm Velocity  

Response Time 

Low Alarm. LAL 

Alarm Set Point 

Alarm Set Point 

Set Point 

Mode 

Icon When Enabled 

Display LCD 

Go to Cross 

   ____________ 

Increment 

Decrement  

Response time  

Sec. Area Mode   

Response Time 

Response Time

Display Cross   

Go to Atmos. 

   ____________ 

Increment 

Decrement  

Sectional Area Sq In  Press. Mode 

 

Cross Sec. Area 

Cross Sec. Area

Display AtPr & 

Go to Power 

 

Increment 

Decrement 

mmHg Icon 

Shutoff Mode 

   ____________

 Atmospheric 

Atmospheric 

  

 

 

Pressure 

Pressure

Display Power 

Go to Real Time  Turn On/off 

 

 

Shutoff 

Mode OR Go to   

   ____________

 

   ____________ 

   

Radio Mode 

  

(Wireless)

Display Radio 

Go to Time 

Turn On/off 

 

 

Mode 

Interval 

   

   ____________ 

   ____________ 

  

Transmission 

 

 

Display Time 

Go to 

 

 

 

Interval  

Transmission Unit 

   ____________

 

   ____________

 

   ____________ 

Transmission ID

Display 

Go to Real 

 

 

 

Transmitter Unit ID  Time Mode 

   ____________

 

   ____________

 

   ____________

 

 Operation 

  Press 

Press 

Press 

Press 

SET

-

MODE

°F-°C

SET

-

MODE

°F-°C

SET

-

MODE

°F-°C

SET

-

MODE

°F-°C

 Model No.          Velocity Range FPM(m/s)  Description

HHF1001A 

0 to 5000 (0 to 25.4)   Air velocity/

temperature 

meter, Fixed 

probe

HHF1001R 

0 to 5000 (0 to 25.4)   Air velocity/temp 

meter with remote 

Telescopic Probe

HHF1001A-W  0 to 5000 (0 to 25.4)   Air velocity/temp 

meter,fixed probe 

& wireless

HHF1001R-W  0 to 5000 (0 to 25.4)   Air velocity/temp 

meter, telescopic 

probe & wireless

Recommended 

Straight Pipe 

Length "A " 

T ypical Piping 

Remarks 

Withou t 

V anes 

With 

Va ne s 

15D 

15 D

C

 losed Branch 

20D 

15 D 

Elbow .  Te e, 
Branch Pipe 

25D 

15D 

Elbow , 2 planes 

25D 

15 D 

Long-radius 
bend s 

30D 
25D 

15D 
15D 

Elbow 
Long-radius 
bend s 

40D 
35D 

15D 
15D 

Elbow 
Long-radius 
bend s 

20D 

15 D

C

 ontracting Pip e 

40D 

20D 

Expanding Pipe 

Recommend Mete r 

Be Installed 

Upstream 

Regulating, 
reducing valves 
Ball, check valves 
Shut-of f valves 

aried Section 

Va 

lves 

Fittings in 

 Tw 

o Planes 

All Fittings in Same Plan 

Note: Straight pipe length on the downstram side to be 5 pipe diameters minimum . 

Note: D – Pipe internal diameter