background image

5. K/J T1 Input Thermocouple Type Select 

(Main display) 

The K/J key selects the T1 thermocouple type, when the 

main  display  is  showing  T1.  When  the  thermometer  is 

turned  on,  it  is  set  to  the  type  selected  when  the  ther-

mometer was last turned off. 

6. MIN MAX with Time record mode 

(Main display) 

Press MIN MAX key to enter the MIN MAX Recording 

mode, (displays the Maximum reading with time, Mini-

mum  reading  with  time  and  Average  reading  stored  in 

record mode). In the this mode the automatic power-off 

feature  is  disabled  and 

key,  °C/°F  key,  REL  key, 

SET key, Hi/Lo Limits key, K/J, C.J. TEMP and Recall 

keys are disabled. The beeper emits a tone when a new 

minimum  or  maximum  value  is  recorded.  Present  tem-

perature reading displayed on second display. Press MIN 

MAX  key  to  cycle  through  the  MAX,  MIN  and  AVG 

readings. If an overload is recorded, the averaging func-

tion is stopped. In this mode, press the HOLD key to stop 

the recording of readings, all values are held, press again 

to restart recording. 

To prevent accidental loss of MIN, MAX and AVG data, 

this mode can only be cancelled by pressing and holding 

the  “MIN/MAX”  key  for 2  seconds.  All  recorded  read-

ings are erased. 

7. REL Relative mode (Main display) 

Press REL key to enter the Relative mode, zero the dis-

play, and store the displayed reading as a reference value 

and  annunciator  REL  is  displayed.  Present  temperature 

reading  displayed  on  second  display.  Press  REL  key 

again  to  exit  the  relative  mode.  The  relative  value  can 

also be entered by the user. (See “SET mode” later in this 

manual.)  When  the  desired  Relative  value  has  been  en-

tered,  press  REL  key  to  enter  the  Relative  mode,  press 

SET key use set Relative value as a reference value. Press 

REL key again to exit the relative mode. In the Relative 

mode, the value (can not >±3000.0 counts) shown on the 

LCD  is  always  the  difference  between  the  stored  refer-

ence and the present reading. 

8. Sec. Minu. Selecting the Time scale 

Press  this  key  to  display  the  elapsed  time  on  the  third 

display in either hours and minutes or minutes and sec-

onds.  When  the  thermometer  is  turned  on,  it  is  set  to 

seconds. To change the time scale, press sec. Minus. key. 

Maximum  elapsed  time  reading  is  100  hours.  If  100 

hours is exceeded, the elapsed time resets to zero. 

9. SET mode (Relative value set, Time set and 

Hi/Lo Limits value set) 

9.1

  Press  the  “SET”  key  to  enter  relative  values  SET 

mode (Press “ENTER” key   to  skip  setting  relative 

value).  =  =  =  =.=  is  displayed  on  the  main  display. 

The  relative  value  is  entered  via  overlay  numbers, 

press overlay “ENTER” key to store the relative value, 

and advance to elapsed time set mode. 

9.2

  Elapsed  time  set  mode,  (Press  ENTER  key  to  skip 

Elapsed time set mode) =.= = = : = = is displayed in 

second and third display. Time (hours, minutes, sec-

onds)  value  is  entered  via  overlay  numbers,  press 

overlay  “ENTER”  key  to  store  time  value.  Elapsed 

time starts from set time value. 

9.3

  Hi  Limit  value  set  mode, 

  is  displayed  (Press 

“ENTER” key to skip Hi Limit value set mode), = = = 

=.=  is  displayed  in  main  display,  Hi  Limit  value  is 

entered  via  overlay  numbers,  then  press  “ENTER” 

key to store the Hi Limit value. = = = =.= is displayed 

in main display, Lo Limit value is entered via overlay 

numbers,  then  press  overlay  “ENTER”  key  to  store 

the Lo Limit value and exit SET mode. 

9.4

 When the thermometer is turned on it uses the relative 

value and Hi/Lo Limits values that were entered when 

thermometer was last in use. 

10. Relative value Recall display mode. 

Press REL RECALL key to display the Relative set value 

on second display. Press REL RECALL key again to exit 

this mode. 

11. Hi/Lo Limits value Recall display mode 

Press  Hi/Lo  RECALL  key  to  display  the  Hi  set  Limit 

value on second display. Press Hi/Lo RECALL key again 

to  display  the  Lo  Limits  set  value  on  second  display. 

Press Hi/Lo RECALL key again to exit this mode. 

12. Hi/Lo Limits mode (Main display) 

Press “Hi/Lo Limits” key to enter the Hi/Lo Limits com-

parative  mode, 

  is  displayed.  When  the  input  tem-

perature  value  exceeds  the  Hi  or  Lo  Limits  value,  the 

beeper  emits  a  continuous  or  pulsed  tone.  Press  “Hi/Lo 

Limits” key again to exit the Hi/Lo Limits mode. 

OPERATOR MAINTENANCE 

WARNING 

To avoid possible electrical shock, disconnect the 

thermocouple connectors from the thermometer before 

removing the cover.

Battery Replacement 

1.  Power  is  supplied  by  4pcs  1.5V  (AAA  SIZE)  UM-4 

R03. 

2. The “ ” appears on the LCD display when replace-

ment is needed. To replace battery remove screw from 

back of meter and lift off the battery cover. 

3. Remove the battery from battery contacts and replace. 

4. When not in use for long periods remove battery. 

5.  Do  not  store  in  locations  with  high  temperatures,  or 

high humidity. 

Cleaning 

Periodically wipe the case with a damp cloth and deter-

gent, do not use abrasives or solvents. 

*Software Operation manual is on the Software disk. 

Where Do I Find Everything I Need for

Process Measurement and Control?

OMEGA…Of Course!

Shop online at omega.com

TEMPERATURE

Thermocouple, RTD & Thermistor Probes, 

Connectors, Panels & Assemblies

Wire: Thermocouple, RTD & Thermistor

Calibrators & Ice Point References

Recorders, Controllers & Process Monitors

Infrared Pyrometers

PRESSURE, STRAIN AND FORCE

Transducers & Strain Gages

Load Cells & Pressure Gages

Displacement Transducers

Instrumentation & Accessories

FLOW/LEVEL

Rotameters, Gas Mass Flowmeters & Flow Computers

Air Velocity Indicators

Turbine/Paddlewheel Systems

Totalizers & Batch Controllers

pH/CONDUCTIVITY

pH Electrodes, Testers & Accessories

Benchtop/Laboratory Meters

Controllers, Calibrators, Simulators & Pumps

Industrial pH & Conductivity Equipment

DATA ACQUISITION

Data Acquisition & Engineering Software

Communications-Based Acquisition Systems

Plug-in Cards for Apple, IBM & Compatibles

Data Logging Systems

Recorders, Printers & Plotters

HEATERS

Heating Cable

Cartridge & Strip Heaters

Immersion & Band Heaters

Flexible Heaters

Laboratory Heaters

ENVIRONMENTAL MONITORING AND CONTROL

Metering & Control Instrumentation

Refractometers

Pumps & Tubing

Air, Soil & Water Monitors

Industrial Water & Wastewater Treatment

pH, Conductivity & Dissolved Oxygen Instruments

WARRANTY/DISCLAIMER

OMEGA  ENGINEERING,  INC.  warrants  this  unit  to  be  free  of  defects  in

materials  and  workmanship  for  a  period  of 

13  months

from  date  of

purchase.  OMEGA’s  WARRANTY  adds  an  additional  one  (1)  month  grace

period  to  the  normal 

one  (1)  year  product  warranty

to  cover  handling

and shipping time. This ensures that OMEGA’s customers receive maximum

coverage on each product. 

If  the  unit  malfunctions,  it  must  be  returned  to  the  factory  for  evaluation.

OMEGA’s Customer Service Department will issue an Authorized Return (AR)

number immediately upon phone or written request. Upon examination by

OMEGA, if the unit is found to be defective, it will be repaired or replaced at

no  charge.  OMEGA’s  WARRANTY  does  not  apply  to  defects  resulting  from

any  action  of  the  purchaser,  including  but  not  limited  to  mishandling,

improper interfacing, operation outside of design limits, improper repair, or

unauthorized  modification.  This  WARRANTY  is  VOID  if  the  unit  shows

evidence of having been tampered with or shows evidence of having been

damaged  as  a  result  of  excessive  corrosion;  or  current,  heat,  moisture  or

vibration; improper specification; misapplication; misuse or other operating

conditions  outside  of  OMEGA’s  control.  Components  in  which  wear  is  not

warranted, include but are not limited to contact points, fuses, and triacs.

OMEGA  is  pleased  to  offer  suggestions  on  the  use  of  its  various

products. However, OMEGA neither assumes responsibility for any

omissions  or  errors  nor  assumes  liability  for  any  damages  that

result from the use of its products in accordance with information

provided  by  OMEGA,  either  verbal  or  written.  OMEGA  warrants

only  that  the  parts  manufactured  by  the  company  will  be  as

specified  and  free  of  defects.  OMEGA  MAKES  NO  OTHER

WARRANTIES  OR  REPRESENTATIONS  OF  ANY  KIND

WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE,

AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF

MERCHANTABILITY  AND  FITNESS  FOR  A  PARTICULAR  PURPOSE

ARE  HEREBY  DISCLAIMED.  LIMITATION  OF  LIABILITY:  The

remedies of purchaser set forth herein are exclusive, and the total

liability  of  OMEGA  with  respect  to  this  order,  whether  based  on

contract,  warranty,  negligence,  indemnification,  strict  liability  or

otherwise, shall not exceed the purchase price of the component

upon  which  liability  is  based.  In  no  event  shall  OMEGA  be  liable

for consequential, incidental or special damages.

CONDITIONS:  Equipment sold by OMEGA is not intended to be used, nor

shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in

or with any nuclear installation or activity; or (2) in medical applications or

used  on  humans.  Should  any  Product(s)  be  used  in  or  with  any  nuclear

installation or activity, medical application, used on humans, or misused in

any  way,  OMEGA  assumes  no  responsibility  as  set  forth  in  our  basic

WARRANTY / DISCLAIMER  language,  and,  additionally,  purchaser  will

indemnify OMEGA and hold OMEGA harmless from any liability or damage

whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS / INQUIRIES

Direct  all  warranty  and  repair  requests/inquiries  to  the  OMEGA  Customer

Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA,

PURCHASER  MUST  OBTAIN  AN  AUTHORIZED  RETURN  (AR)  NUMBER

FROM  OMEGA’S  CUSTOMER  SERVICE  DEPARTMENT  (IN  ORDER  TO

AVOID  PROCESSING  DELAYS).  The  assigned  AR  number  should  then  be

marked on the outside of the return package and on any correspondence.

The  purchaser  is  responsible  for  shipping  charges,  freight,  insurance  and

proper packaging to prevent breakage in transit.  

FOR 

WARRANTY

RETURNS, please

have the following information

available BEFORE contacting

OMEGA:

1. Purchase Order number under

which the product was

PURCHASED,

2. Model and serial number of the

product under warranty, and

3. Repair instructions and/or specific 

problems relative to the product.

FOR 

NON-WARRANTY

REPAIRS,

consult OMEGA for current repair

charges. Have the following

information available BEFORE

contacting OMEGA:

1. Purchase Order number to cover

the COST of the repair,

2. Model and serial number of the

product, and

3. Repair instructions and/or specific

problems relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever

an improvement is possible. This affords our customers the latest in

technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2011 OMEGA ENGINEERING, INC. All rights reserved. This

document may not be copied, photocopied, reproduced, translated, or

reduced to any electronic medium or machine-readable form, in whole or

in part, without the prior written consent of OMEGA ENGINEERING, INC.

Reviews: