Adjustment key of thermocouple 0°C:
Main display shows (T1 or T2) readings, use and main display
(T1 or T2) thermocouple of the same type,
imports input of main display 0°C (32°F) of
standard temperature value, adjust the cold
junction compensation of third display, enables
main display to show 0°C (32°F), then press
key finish adjusting. The adjusting range is about
0°C ~50°C (32°F~122°F) of third display cold
junction compensation, exceed range show Err
and unable to adjust, need to press key to
give up and left. The cold junction compensation
adjusting (cancelled) in power off, while starting
the meter again will recover the normal cold
junction compensation.
Press once:
Enter adjust manually and third
display shows the cold junction compensation,
enable main display shows 0°C (32°F).
>
Press to move right one-figure that adjusted.
>
Press to move left one-figure that adjusted.
>
Press to add one-figure upward that adjusted.
>
Press to minus one downward that adjusted.
Press for 2 seconds:
Adjust once automatically the
cold junction compensation of the third display,
Make main display shows 0°C (32°F). If need to
adjust again, enter again after press key to
leave.