Sensor Replacement.

  The sensor ordinarily does not need 

replacement unless it is electrically damaged.  If replace-

ment is necessary, unthread the sensor by hand.  Thread 

the replacement sensor in and tighten by hand.

MAINTENANCE and REPAIR

Rotor Replacement.

  There is only one moving part to this 

meter.  The bearings are made of ruby, which rarely wears 

out or needs replacement unless they have been physically 

damaged by severe shock.  The shaft is integrally molded 

into the rotor, and shaft and rotor are replaced as one part. 

(You may wish to replace the bearings, using the bearing 

removal  tool,  while  the  meter  is  disassembled  for  rotor 

replacement).  To replace the rotor, disconnect the meter 

and remove the four screws that hold the cover in place.  

Lift the cover and remove the rotor (see parts diagram).

When putting in the new rotor, be sure that the ends of 
the shaft are in both bearings before tightening the cover. 
The rotor can be easily dropped into the bottom bearing. 
Starting the shaft into the upper bearing requires a bit of 
care. It is easier if the rotor is spinning, which can be done 
by lightly blowing into a port. When the upper bearing plate 
drops into place, hold it down and check for free spinning 
(by blowing lightly) before replacing the cover. Check that 
the O-ring is in its seat on the bearing plate before replacing 
the cover. Replace the cover, insert the four cap screws 
and tighten.

(BLACK) Power (-)
(WHITE) Signal
(RED) Power (+) 6-24 Vdc

SES SENSOR CONNECTION

Reconnect the sensor according to the diagram below.

Summary of Contents for FPR300 Series

Page 1: ...FPR310 FPR300 FPR300 310 SERIES Low Flow Meter User sGuide Shop online at omega com email Info omega com For latest product manuals omegamanual Info...

Page 2: ...78280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056...

Page 3: ...bearings to allow for very low minimum flow rates and superior life With a body material of polypropylene the FPR300 is an economical choice for metering water or low corrosion fluids The lens cover i...

Page 4: ...ound Signal DC Voltage Ground Figure 1 Figure 2 Input Designed for Current Sinking NPN Devices Input Designed for Current Sourcing PNP Devices 2 2K Ohm Pull up Resistor NPN Device NPN Device INSTALLAT...

Page 5: ...e rotor disconnect the meter and remove the four screws that hold the cover in place Lift the cover and remove the rotor see parts diagram When putting in the new rotor be sure that the ends of the sh...

Page 6: ...29 11129 Kynar ceramic 6 magnet high res 11132 11132 Kynar cabide 6 magnet high res 11130 11130 5 O Ring EPDM 25081 TFE coated Viton 31403 6 Cover Before 5 05 After 5 05 Before 5 05 After 5 05 Polypro...

Page 7: ...otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 8: ...rbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Conductivity...

Reviews: