background image

2

THE FOLLOWING CONFIGURATIONS ARE POSSIBLE:

1)  Panel Mounting, using hex nuts at rear of meter.
2)  Full In-line Mounting (not available for meters with  valve options.)
3)  Partial In-line Mountings, i.e. inlet at bottom, outlet at top rear, or inlet at bottom 

rear and outlet at top center. 

To Exchange Scales: 

Select Option A, B, or C

A. NO VALVE OPTION (Styles 2, 4, 14)

1)

Using a 1/2 open box wrench or 5/16" hex wrench remove rear fittings (Female 
Adaptor (9) for the panel mounting option or Plug Seal (3) for the in-line mounting
option).

2)

Remove Back Plate (2) and Scale (7) by sliding both to either side.

3)

Holding the desired scale align its holes with those in the Back Plate. Slide both 
behind the bracket (5).

4)

Insert but do not completely tighten fittings.

5)

Recheck Back Plate (2) alignment.

6)

Tighten fittings.

B. VALVE OPTION (Styles 2,4)

1)

Using a 1/2 open box wrench remove top Female Adaptor (9).

2)

Loosen Valve Cartridge (11) allowing the lower female (12) to be unscrewed.   

Do not remove Cartridge from the meter.

3)

Remove Back Plate (2) and Scale (7) by sliding both to either side.

4)

Holding the desired scale align its holes with those in the Back Plate. Slide both 
behind the bracket (5).

5)

Insert but do not completely tighten top Female Adaptor (9).

6)

Insert but do not completely tighten lower Female Adaptor (12) (it has an extra 
orifice and O-ring on it), while keeping Valve Cartridge (11) in place.

7)

Screw Valve Cartridge (11) into lower Female Adaptor (12).

8)

Recheck Back Plate (2) alignment.

9)

Tighten both Female Adaptors (9), (12).

10) Tighten Valve Cartridge (11). 

Do not over tighten it.

C. VALVE OPTION (Style 14)

1)

Use a 1/16" hex wrench to loosen 2 set screws on Knob (15). 

2)

Using a 1/2" open box wrench, remove Valve Cartridge (13) and top Female 
Adaptor (9). 

3)

Remove Back Plate (2) and Scale (7) by sliding them to either side.

4)

Holding the desired scale align its holes with those in the Back Plate slide them 
behind Brackets (5).

5)

Insert but do not completely tighten top Female Adaptor (9).

6)

Insert but do not completely tighten Valve cartridge (13).

7)

Recheck Back Plate (2) alignment.

8)

Tighten Female Adaptor (9) and Valve cartridge (13).

9)

Secure Knob, tighten its set screws.

Summary of Contents for FL-2500

Page 1: ...FL 2500 Acrylic Flow Meters www omega com e mail info omega com User sGuide Shop online at...

Page 2: ...e Tel 33 0 1 61 37 29 00 FAX 33 0 1 30 57 54 27 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Tel 49 0 7056 9398 0 FAX 49 0 7056...

Page 3: ...Carbide Carboloy silver colored FITTINGS Brass or 316 Stainless Steel O RINGS FKM is standard Buna EPR are optionally available VALVES Built in needle valves are available MAXIMUM SERVICE TEMPERATURE...

Page 4: ...move Cartridge from the meter 3 Remove Back Plate 2 and Scale 7 by sliding both to either side 4 Holding the desired scale align its holes with those in the Back Plate Slide both behind the bracket 5...

Page 5: ...he float should be handled with appropriate tweezers Do not touch it with fingers Store the float in a clean bag or box free of grease A NO VALVE OPTION Styles 2 4 14 Unscrew top Plug Seal 3 using a 5...

Page 6: ...ottom Plug Seal 3 until it is flush with Bracket 5 Proceed to steps 1 3 C VALVE OPTION Style 14 Insert the plastic Y shape Float Stop 14 from the wide bore using an appropriate non metal stick push th...

Page 7: ...5 In Line Version Styles 2 4 14 Valve Option Version Styles 2 4...

Page 8: ...el 4 4 O ring FKM 5 2 Bracket Structural aluminum 6 1 Float Glass or 316 Stainless Steel or Carboloy 7 1 Scale Film 8 2 Bumping Spring 316 Stainless Steel 9 1 or 2 Female Adaptor Brass or 316 Stainles...

Page 9: ...helene 3 2 Seal Plug Brass or 316 stainless steel 4 3 O ring FKM 5 2 Bracket Structural aluminum 6 1 Float 316 Stainless Steel 7 1 Scale Film 8 1 Bumping Spring 316 Stainless Steel 9 1 Female Adapter...

Page 10: ...8 DIMENSIONAL DRAWINGS...

Page 11: ...se shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold by OME...

Page 12: ...Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH...