FDT-25 Handheld Ultrasonic Flowmeter
Other menu windows such as M88 have no functions, or functions were cancelled because they are
not applied to this version of the software.
§2.6 Steps to Configure the Parameters
The following parameters need to be configured for a proper measurement:
(1) Pipe outer diameter
(2) Pipe wall thickness
(3) Pipe materials (for non-standard pipe materials*, the sound speed for the material must be
configured too)
*Standard pipe materials and standard liquids refer to those with the sound parameters that have
already been programmed into software of the flow meter, therefore there is no need to configure
them
(4) Liner material and its sound speed and thickness, if there is any liner.
(5) Liquid type (for non-standard liquids, the sound speed of the liquid is also needed)
(6) Transducer type adapted to the flow meter. Generally the Standard M1 or the Frame M-sized
transducers will be the selected option.
(7) Transducer mounting methods (the V-method or Z-method is the common option)
(8) Check up the Space displayed on M25 and install the transducers accordingly.
For standard pipe materials and standard liquids, the following detailed step-by-step setup is recommended.
(1) Press keys
and then press
(2) Press key
press ENT key.
1 1 to enter M11 window to input the digits for the pipe outer diameter,
ey.
enter M12 window to input the digits for the pipe outer diameter and then
(3) Press key
to enter M14 window, and press ENT key to enter the option selection mode.
Use keys
(4) Press key
keys
and
to scroll up and down to the intended pipe material, and then press ENT key.
nter M16 window, press ENT key to enter the option selection mode, use
roll up and down to the liner material, and then press ENT key. Select “No
Liner”, if there is no liner.
(5) Press key
to enter M20 window, press
key to enter the option selection mode, use
keys
and
to scroll up and down to the proper liquid, and then press ENT key.
(6) Press key
to enter M23 window, press
key to enter the option selection mode, use
keys
and
to scroll up and down to the proper transducer type, and then press ENT key.
(7) Press key
to enter M24 window, press
key to enter the option selection mode, use
keys
and
to scroll up and down to the proper transducer mounting method, and then press
ENT key.
(8) Press key
to enter M24 window to install the transducers on the pipe, and then press
key to go to M01 for the results.