24
Configuration / operation
10
Section 10.5 - Advanced Functions
The unit gives access to a number of advanced functions which can be reached by
answering “Yes” to the point “adv.set”.
Section 10.6 - Display setup
Here you can adjust the brightness contrast and the backlight. Setup of TAG number
with 6 alphanumerics. Selection of functional readout in line 3 of the display - choose
between readout of analogue output or TAG number.
Section 10.7 - Two-point process calibration
The unit can be process-calibrated in 2 points to fit a given input signal. A low input
signal (not necessarily 0%) is applied and the actual value is entered via DRSL-
DISPLAY. Then a high signal (not necessarily 100%) is applied and the actual value
is entered via DRSL-DISPLAY. If you accept to use the calibration, the unit will work
according to this new adjustment. If you later reject this menu point or choose another
type of input signal the unit will return to factory calibration.
Section 10.4 - Signal and sensor error info (no display)
Status of the unit can also be read from the red/green LED in the front of the device.
Green flashing LED 13 Hz indicates normal operation.
Green flashing LED 1 Hz indicates sensor error.
Steady green LED indicates internal error.
Steady red LED indicates fatal error.