background image

Ranges and Resolution

See table below. Resolution is fixed as indicated in table. 

Accuracy

Accuracy includes linearity, hysteresis, repeatability 
±0.25% of full scale ±1 least significant digit 
Sensor hysteresis: 

±0.015% FS, included in accuracy 

Sensor repeatability: 

±0.01% FS, included in accuracy

Display

4 readings per second nominal display update rate 
4 digit LCD, 0.5” H and 5 character 0.25” H alphanumeric 
Alarm 1 and Alarm 2 LCD indicators and bi-color (red/green) 
LEDs

Controls & Functions

SEL  Select and display alarm trip points 
TEST  Alarm acknowledge, or toggle alarms when in test mode 

s

  Increase alarm setpoint when in setpoint adjust mode 

t

  Decrease alarm setpoint when in setpoint adjust mode 

Multi-level pass code protection for set-up and calibration

Alarm Relay Outputs

Programmable dual form C (SPDT) relay contacts for HI/LO, 
HI/HI, LO/LO, normal or reverse acting with 1% deadband, or 
adjustable trip and reset points for each relay, manual or auto 
acknowledge.
1A/24VDC, 0.5A/115VAC, non-inductive 
120 milliseconds typical response time

4-20 mA Retransmission Output

Updated approximately 16 times per second
Current output, 4-20 mA DC 
Output drive (compliance) determined by power source 
6,553 counts over sensor range

Calibration

Non-interactive, ±10% of range 
All pressure and absolute models: zero, midpoint, span 
All vacuum models:  –span, –midpoint, zero 
Vacuum/pressure models:  –span, zero, +midpoint, +span 
±15 psi models: –span, –midpoint, zero, +midpoint, +span

¼" NPT

2.88"

3.38" Standard 

3.5" NEMA 4X

0.75"

1.65"

X X X X X

88888

2

3 ft long, 22 AWG cables 

with stripped and tinned 

wire ends 

4 conductor and 

6 conductor
Allow 1" wire clearance

Turn at hex 

fitting only

SEL TEST

t

s

ALARM 1

ALARM 2

Alarm1

Alarm 2

Select 

Test

Up 

Down

Alarm 1 

status 

LED

Numeric 

display

Alpha-

numeric 

display

Alarm 2 

status 

LED

DPG1000DAR Series

M3363/0818

Digital Pressure Gauges with Alarms

 and 4-20 mA Output

INSTRUCTION

SHEET

Shop online at omega.com  e-mail: [email protected] 
For latest product manuals: www.omegamanual.info

TM

Power

8 to 24 VAC 50/60 Hz or 9 to 32 VDC 
Gauge is on when power is on 
Designed for continuous operation 
1.0 watt maximum power consumption

Weight

9.5 ounces (approx.), shipping wt. 1 pound (approx.) 

Housing

Extruded aluminum case, epoxy powder coated, ABS/ polycar-
bonate bezel, front and rear gaskets, polycarbonate label

Connection, Material, Media Compatibility

1/4” NPT male fitting 
All wetted parts are 316L stainless steel

Overpressure, Burst, Vacuum

Ranges using 3000 psig sensor:  5000 psig overpressure 
Ranges using 5000 psig sensor:  7500 psig overpressure 
All others: 

2 X pressure range

3000 psi, 5000 psi, and 4 digit ranges 112.5% full scale out-
of-range display: 1– – – or I –.–.–.–
Under-range display (non-vacuum sensors): –Err
4 X sensor burst pressure rating, or 10,000 psi, whichever is 
less
Vacuum service: 15 psia, ±15 psig, 15 psig, 30 psia,  
 

100 psig, 100 psia, 200 psig sensors

Environmental

Storage temperature: 

–40 to 203°F (–40 to 95°C)

Operating temperature: 

–4 to 185°F (–20 to 85°C)

Compensated temperature: 

32 to 158°F (0 to 70°C)

Dual Alarms, 4-20 mA Output

Range

DPG1000DAR-30V100

–30.0 inHg to 100.0 psig

DPG1000DAR-30INHGVAC

 

0.00-30.00 InHg Vac

DPG1000DAR-05G

 

0.00-5.000 psig

DPG1000DAR-15A 

 

0.00-15.00 psia

DPG1000DAR-15G

 

0.00-15.00 psig

DPG1000DAR-30A

 

0.0-30.00 psia

DPG1000DAR-30G

 

0.0-30.00 psig

DPG1000DAR-60G

 

0.0-60.00 psig

DPG1000DAR-100A

 

0.0-100.0 psia

DPG1000DAR-100G

 

0.0-100.0 psig

DPG1000DAR-300G 

 

0-300.0 psig

DPG1000DAR-500G

 

0-500.0 psig

DPG1000DAR-1KG

 

0-1000 psig

DPG1000DAR-2KG

 

0-2000 psig

DPG1000DAR-3KG

 

0-3000 psig

DPG1000DAR-5KG

 

0-5000 psig

Alarm Setting 

(All are field configurable)

Alarm Type

-1N

 (default if not specified)

HI/LO Normal action

-2N

HI/HI Normal action

-3N

LO/LO Normal action

-1R

HI/LO Reverse action

-2R

HI/HI Reverse action

-3R

LO/LO Reverse action

WARNING:

 This product can expose you to chemicals including 

lead, nickel and chromium, which are known to the State of 

California to cause cancer or birth defects or other reproductive 

harm. For more information go to www.P65Warnings.ca.gov