The instrument is a panel mount device protected in accordance

with EN61010-1 (Safety requirements for electrical equipment

for measurement, control, and laboratory standard). Remember

that the unit has no power-on switch. Building installation should

include a switch or circuit-breaker that must be compliant to IEC

947-1 and 947-3.

SAFETY:

• Do not exceed voltage rating on the label located on the top

of the instrument housing.

• Always disconnect power before changing signal and power

connections.

• Do not use this instrument on a work bench without its case

for safety reasons.

• Do not operate this instrument in flammable or explosive

atmospheres.

• Do not expose this instrument to rain or moisture.

EMC:

• Whenever EMC is an issue, always use shielded cables.

• Never run signal and power wires in the same conduit.

• Use signal wire connections with twisted-pair cables.

• Install Ferrite Bead(s) on signal wire close to the instrument if

EMC problems persist.

To Select Operating Mode

(Rate or Total):

1.

Press  MENU  until the

display shows 

Func

.

2.

Press   SET to show

the current mode.

3.

Press   SET to

select between 

rAtE

and 

totAL

.

4.

Press MENU to store

the mode into volatile

memory.

To store the operating mode (or any value)

into non-volatile memory:

Press MENU until the meter displays 

noStor

.

Press  /SET once to display 

StorE

.

Press MENU again.

The value is stored in non-volatile memory.

3-Wire Sensor

Open Collector

(PNP)

2-Wire Contact

Closure

2-Wire

Magnetic

Pickup

Using the Menus

Menu

Menu

Prompt is starting

Prompt

Item

point:

Func

Function

to select rate or totalize

mode.

SCALE

Scale Factor

to select scale factor

as a divider or multiplier

C

S

.

 

A

and to enter scale 

A/SC

factor value.

OFFSEt

Offset Value

to select offset value.

Reading on meter =

input x scale + offset

dEC Pt

Decimal Point to select position of

decimal point in the

digital display.

SP LO

Setpoints

to enter setpoints. 

Low and

Shows only if 

SP HI

High

the relay option 

board is installed.

COnFIG

Configuration to program internal

software switches to 

specify analog,

output, RS-232 input 

or alarm conditions

An LO

Analog Low

to obtain reading for

low analog output.

An HI

Analog High

to obtain reading for

high analog output.

NoStor

No Store

to save to the volatile

(active) memory

StorE

Store

to save to the non-

volatile (permanent) 

memory.

RECALL

/ADV

MENU

RESET

/SET

Func

RECALL

/ADV

MENU RESET

/SET

rate

totAL

RECALL

/ADV

MENU RESET

/SET

RECALL

/ADV

MENU RESET

/SET

store

Wiring Examples

EXC

RTN

1

2

3

4

5

6

OPEN

OPEN

OPEN

CLOSED

OPEN

CLOSED

OPEN

OPE

N

CLOSED

OPEN

OPEN

OPEN

POSITIVE THRESHOLD

1K PULL DOWN RESISTOR TO RTN
243K PULL UP RESISTOR TO +EXC

S1

1
2
3
4
5
6
7
8
9
10

TB2

12.4 V

OUTPUT

EXC OUT

SIG IN

SIG RTN

RST IN

EXT GATE

BATTERY

EXCITATION = 12.4V

1

2

3

4

5

6

12.4V EXC

OPEN

OPEN

OPEN

CLOSED

OPEN

CLOSE

D

CLOSED

OPE

N

CLOSED

OPEN

OPEN

OPEN

HIGH LEVEL SIGNAL

POSITIVE THRESHOLD

LOW PASS FILTER

1K PULL DOWN RESISTOR

S1

1
2
3
4
5
6
7
8
9
10

TB2

EXC OUT

SIG IN

SIG RTN

RST IN

EXT GATE

BATTERY

243K PULL UP RESISTOR TO +EXC

1

2

3

4

5

6

EXCITATION= 5.0V

CLOSED

OPEN

CLOSED

OPEN

OPEN

CLOSE

D

OPEN

OPE

N

OPEN

OPEN

OPEN

OPEN

LOW LEVEL SIGNAL

ZERO THRESHOLD

NO FILTER

243K PULL-UP RESISTOR TO +EXC

S1

1
2
3
4
5
6
7
8
9
10

TB2

EXC OUT

SIG IN

SIG RTN

RST IN

EXT GATE

BATTERY

AC Powered Unit Connections

CAUTION: Do not connect AC power to your device until

you have completed all input and output connections. This

device must only be installed by a specially trained

electrician with corresponding qualifications. Failure to

follow all instructions and warnings may result in injury!

This device is marked with the international “Caution, risk of

danger” symbol.

External Fuse Required:

Time-delay, UL 248-14 listed

125 mA (115 Vac line)

63 mA (230 Vac line)

Time-lag, IEC 127-2 recognized

100 mA (115 Vac line)

50 mA (230 Vac line)

For complete wiring information on all setup

options, please refer to the Operator’s Manual.

Using this Quick Reference

Use this Quick Reference with your controller for information

on wiring and to make changes to the operating mode, scale

factor, and decimal point position.  For detailed instructions,

refer to the appropriate section in the Reference Manual.

Safety Consideration

Mount the Unit

1. Cut a panel opening using the dimensions shown to the right.

2. Position the unit in the opening, making sure the front bezel is 

flush with the panel.

3. Install retaining clips on both sides of the meter and tighten 

against the panel.

Wiring

1. Remove the panel at the back

of the unit.

2. Locate the TB1 connector.

3. Insert the correct wire in each 

terminal as shown in the

following figure and tighten the lockdown screws.

4. Tug gently on the wires to verify the connections.