Operating Instructions
3-8
Backlight and Backlight Timeout
The instrument includes an LED backlight feature to ensure measurement data can be easily
read in low-light conditions. To activate the backlight, press
.
Once the backlight is activated, it will automatically turn off after 30 seconds if no key is
pressed to preserve battery life. To disable the backlight timeout feature, press
(1.5s)
. The
backlight will flash to indicate the timeout feature has been disabled. To re-enable the
backlight timeout feature, turn the backlight off then on by pressing
twice.
Operating Modes
The instrument has five (5) operating modes for channel one, “Source” including manual
operation explained in
Figure 8 below
. The operating modes are Manual, Fast ramp, slow
ramp, step and transfer.
O
PERATING
M
ODE
D
ISPLAY
I
NDICATOR
D
ESCRIPTION
Manual
blinking
“__ __ __ __”
Instrument operates by outputting a voltage that
corresponds to the set temperature or
millivoltage.
Fast Ramp
Instrument Ramps the output temperature or
millivoltage from 0% of span to 100% of span and
back to 0% of span in 20 seconds. This repeats
until stopped. Any other key press will stop the
output at the existing value (temperature or
millivoltage) and clear the fast ramp setting.
Slow Ramp
Instrument Ramps the output temperature or
millivoltage from 0% of span to 100% of span and
back to 0% of span in 120 seconds. This repeats
until stopped. Any other key press will stop the
output at the existing value (temperature or
millivoltage) and clear the slow ramp setting.
Step
Instrument Ramps the output temperature or
millivoltage from 0% of span to 100% of span and
back to 0% of span stepping at 10% increments,
dwelling 5 seconds at each step. This repeats
until stopped. Any other key press will stop the
output at the existing value (temperature or
millivoltage) and clear the Step ramp setting.
Transfer
Instrument sets the source output (temperature
or millivoltage) equal to the value on the “Read”
channel, (channel 2). This is an offset corrected
value. SOURCE output = READ v CJC
voltage. This mode would be used for
troubleshooting systems and readouts. The
“Transfer” icon will blink and illuminate if selected.
Figure 8: Operating Modes
Summary of Contents for CL940A
Page 39: ...Appendices B iii Thermocouple Type K ...
Page 40: ...Appendices B iv ...
Page 41: ...Appendices B v Thermocouple Type J ...
Page 42: ...Appendices B vi ...
Page 43: ...Appendices B vii Thermocouple Type T ...
Page 44: ...Appendices B viii ...
Page 45: ...Appendices B ix Thermocouple Type E ...
Page 46: ...Appendices B x ...
Page 48: ...Appendices B xii ...
Page 50: ...Appendices B xiv ...
Page 51: ...Appendices B xv Thermocouple Type R ...
Page 52: ...Appendices B xvi ...
Page 53: ...Appendices B xvii Thermocouple Type S ...
Page 54: ...Appendices B xviii ...
Page 55: ...Appendices B xix Thermocouple Type G ...
Page 56: ...Appendices B xx ...
Page 57: ...Appendices B xxi Thermocouple Type C ...
Page 58: ...Appendices B xxii ...
Page 59: ...Appendices B xxiii Thermocouple Type D ...
Page 60: ...Appendices B xxiv ...
Page 61: ...Appendices B xxv Thermocouple Type P ...
Page 62: ...Appendices B xxvi ...
Page 63: ...Appendices B xxvii Thermocouple Type L ...
Page 64: ...Appendices B xxviii ...