7
6. Battery Save Feature
The CL300A incorporates an Auto Shut Off or Battery Save feature that will
automatically turn off the CL300A after a preset amount of time (up to 30
minutes) The factory default setting is 30 minutes, but can be changed or
disabled by completing the following steps:
1) Turn on the CL300A. Closely watch the display during power up.
2) The display will indicate the amount of time the battery save feature is set
for (off to 30 minutes). Immediately adjust the knob to change the time
period or disable the function.
7. Setpoint Feature
The CL300A allows the user to program up to 3 setpoints for each TC type
(including mV). to set the setpoints complete the following procedure:
1) Turn on the CL300A and select the Source Mode.
2) Select the TC type to be simulated and correct units (either °C or °F).
3) Press
to see what is currently stored in each of the 3 set-
points.
4) To store a new value adjust the output to the desired value then press
the key.
5) Repeat steps 3 & 4 until all 3 setpoints are stored.
6) Once the setpoints are stored, press the
key to recall the
newly stored setpoints.
7) Each press of the
key will increment to the next stored set-
point.
Note
: Setpoints are non-volatile and will remain in memory until changed.
8. Min/Max Datalogging
When the CL300A is in the measure mode it will continuously monitor and
store the minimum and maximum values for that particular input. To oper-
ate the datalogging feature use the following procedure:
1) Place the CL300A in the measure mode.
2) Connect a thermocouple (or probe).
Recall
Recall
Set
Recall