Figure 2. Limitation of Maximum Setpoint Temperature at Elevated Ambient
Temperatures
An increase in ambient temperature of 1°C reduces the maximum allowable
target plate temperature by 3.7°C. (An increase of 1°F is accompanied by an
decrease in minimum target plate temperature of 6.7°F).
2.4 Power Connection
Standard (115 Vac~, 50/60 Hz models)
The BB703 comes with a standard North American 3-prong AC power cord. Do
not use any other power cord other than the one provided. This cord provides
the proper grounding and has been safety tested by the proper safety agencies.
International (230 Vac~, 50/60 Hz models)
On 230 Vac~, 50/60Hz models an International style power cord with the proper
color code and approvals is provided with stripped wire ends for connection
to the proper connector used in your country or local area, this connector is not
provided.
Electrical connections and wiring should be performed only by suitably
trained personnel.
Be sure that the line voltage powering your unit does not go above or below 10%
of the rated voltages specified above.
BB703 Installation
2-2
23.9
26.7
29.4
32.2
35.0
37.8
40.6
750
730
710
690
670
650
630
610
590
570
550
398.8
387.8
376.7
365.6
354.4
343.3
332.2
321.1
310.0
298.9
287.8
AMBIENT TEMPERATURE (C)
MAXIMUM ALLOWABL
E
TARGET PLATE TEMPERATURE (F
)
MAXIMUM ALLOWABL
E
TARGET PLATE TEMPERATURE (C
)
AMBIENT TEMPERATURE (F)
75
76.4
85
80
90
95
100
105
OPERATING
REGION
104.0
24.7
40.0
CAUTION:
Summary of Contents for BB703