Ambient Temperature

The target plate of the BB701 can achieve any 

temperature within the specified temperature range 

of -18 to 149°C (0 to 300°F) when being operated 

in normal ambient temperature environments. The 

maximum specified target plate temperature of 149°C 

(300°F) can be achieved over the entire specified 

ambient temperature range. However, lower target 

plate temperatures are more difficult to attain at 

increased ambient temperatures. As long as the 

ambient temperature does not exceed 29.4°C (85°F), 

the target plate will achieve its lower limit temperature 

of -18°C (0°F). The minimum target plate temperature 

the unit can achieve is proportionally worse with 

increased ambient temperature. An increase of 1°C is 

accompanied by an increase in minimum target plate 

temperature of 0.3°C (An increase of 1°F is accompanied 

by an increase in minimum target plate temperature of 

0.5°F).

The Effect of Increased Ambient Temperature  

On Operating Temperatures

Front Panel

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

START HERE

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

2

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

3

ARTWORK/PRODUCT ART/

 DWGS/START HERE ARROW

4

Using This Quick Start Manual

Use this Quick Start Manual with your BB701 

Blackbody Calibrator for easy installation and 

basic operation. For detailed information, refer to 

the User’s Guide (Manual # M3264).

PRECAUTIONS:

•  Follow all safety precautions and operating 

instructions outlined in this quick start and 

accompanying User’s Guide.

•  Never leave your calibrator unattended when in 

use.

• Keep out of reach of children.

• Never touch the target plate when hot.

•  Never place any object within 3 inches of the 

cavity opening when hot.

•  Do not operate in flammable or explosive 

environments.

•  Never operate with a power cord other than the 

one provided with your unit.

•  Remove and or disconnect main power cord 

before attempting any maintenance or fuse 

replacement.

•  Do not connect and or operate this unit to a non-

grounded, non-polarized outlet or power source.

•  Do not connect the serial port or reference probe 

port to equipment with exposed, hazardous, live 

voltages.

General Information

The Model BB701 is a portable, rugged, bench-

top, hot/cold blackbody calibration source with 

a built-in precision PID digital controller. The 

calibrator is used to test and calibrate infrared 

pyrometers. The large 2.5 inch diameter target 

plate has an emissivity of 0.95 and can be set to 

any temperature between –18 to 149°C (0 to 300°F).

Mounting

Mount the unit on a bench, table top or shelf in a 

horizontal position and operate at least ten inches 

from any air obstructions to the fan, front panel, 

rear panel, and top of the unit. Operate the unit in 

an ambient environment between the specified 4.4 

and 43°C (40 to 110°F).

Process Temperature:

This field displays the current temperature of the target 

plate.
Setpoint Temperature:

This field displays the desired target plate temperature. 

Once the target plate reaches this desired temperature, 

both displays will read the same value.
Hot Cycle Indicator:

When this Amber L.E.D. is illuminated, the unit is 

heating up the target plate.
Cold Cycle Indicator:

When this Blue L.E.D. is illuminated, the unit is cooling 

down the target plate.

21.1

+11.8

+7.1

+2.38

0

-2.34

-7.06

-11.2

-13.8

-16.5

-17.8

-19.1

-21.7

26.7

29.4

32.2

AMBIENT TEMPERATURE (

°

C)

MINIMUM A

TT

AINABLE 

TARGET PLA

TE 

TEMPERA

TURE (

°

F)

MINIMUM A

TT

AINABLE 

TARGET PLA

TE 

TEMPERA

TURE (

°

C)

AMBIENT TEMPERATURE (

°

F)

37.8

43.3

70

80

85

90

100

110

TARGET PLATE TEMPERATURE OPERATING REGION

C

AU

TI

ON

: T

AR

GET 

SURFACE M

AY B

E H

O

C

A

U

TI

O

N: 

TA

R GE

T S URFACE

 M

AY

 B

E

 H

O

T

 

BB701

TM

BLACK POINT

Blackbody

Calibrator

00

°

1

°

C

0

°

1

°

C

T

rat r

a

SET OINT TEM ERATURE

ROCESS TEM ERATURE

MO E  EY

TARGET 

ATE

O ER  EY

RAISE  EY

ARAMETER

ACCESS  EY

O ER

S ITCH

HOT CYC E IN ICATOR

CO

 CYC E IN ICATOR

Parameter/Access Key:  

Press to scroll through menu parameters.

  

Raise Key: Press to increase the selected parameter or 

scroll upward in the list of possible settings.

  Lower Key: Press to decrease the selected parameter 

or scroll downward in the list of possible settings.

    Mode Key: 

Press to save settings and exit a menu level.

Back Panel

AIR PURGE NOZZLE

AC POWER INPUT

SERIAL PORT

FAN

FOOTING

REFERENCE PROBE PORT

VENT

AC Power Input:  

The customer connects the power cord to the AC 

Power Input. As a safety precaution, the power 

cord cannot be connected if the fuse compartment 

is open.
Reference Probe Port:  

The reference probe port enables the user to 

monitor the target plate temperature with an 

external instrument. The wires are connected to a 

platinum 3-wire RTD.
Air Purge Nozzle:  

A dry gas source can be connected to this nozzle in 

order to minimize condensation on the target plate 

when operating at low temperatures.
Serial Port:  

The female DB-9 port allows the customer to make 

a 3-wire RS232 interface with the BB701.

Calibrating An IR Pyrometer

 

The 2.5" target plate is a near ideal blackbody 

source. The emissivity of the plate is 0.95. When 

calibrating an IR pyrometer, hold the pyrometer 

perpendicular to the target plate for optimal 

performance. The proper distance between the IR 

pyrometer and the target plate depends on the 

field of view of the pyrometer. If the pyrometer is 

too far away it will scan unwanted surfaces outside 

of the perimeter of the target plate. Holding the 

pyrometer too close could introduce undesirable 

heat to the IR detector of the pyrometer. 

Changing the Temperature Setpoint

 

The BB701’s upper display indicates the blackbody 

target plate temperature known as (PV) Process 

Variable, while the lower display indicates the 

programmed setpoint known as (SV) Setpoint 

Variable. Making changes to the setpoint, units can
be made via the 

 and  

 keys. Holding a key

in, continuously, will cause the setpoint 

temperature to advance more quickly to a desired 

value. Three scanning speeds are provided: slow, 

medium and fast. The minimum and maximum 

setpoints are locked at 0 and 300°F, respectively. 

While these max. and min. settings are changeable, 

it is strongly advised not to adjust these 

parameters.

Reviews: