are made via the
and
keys. Holding a key in, continuously, will cause the
setpoint temperature to advance more quickly to a desired value. Three scanning
speeds are provided: slow, medium and fast. The minimum and maximum setpoints
are locked.
3.5
Changing the Controller Parameter Settings
The BB-4A operates at its optimum performance when left with its factory
parameter settings. The only internal parameter that the operator should need to
change is the engineering units (°F or °C), or serial communications parameters.
Below are two diagrams: a) menu hierarchy with factory default settings
b) programming procedure.
Figure 4. Menu Hierarchy Showing Factory Default Settings
BB-4A Series Blackbody Calibrator
3
Menu 00
Menu 01
Menu 02
Menu 03
Menu 04
Menu 05
Key Lock
SETPOINT
Ac.Cd = 02
Ac.Cd = 03
Ac.Cd = 04
Ac.Cd = 05
Ac.Cd
Gn.o1
113
ALr1
1900
id.no
01
SnSr
c.A
Gr.o2
ALr2
bAUd
12.o.7
Sn.00
rAtE
24
Cy.t1
01
CAL.L
dEC.P
rSEt
122
Cy.t2
CAL.H
FILt
H.Hys
SP.tt
OFF
OUt.1
Ht.P
HyS.1
L.SP.L 212
OUt.2
Alr
C.HyS
L.SCL
CoL.t
nor
HyS.2
U.SP.L 1800
A1.HL
HI
C.SPr
H.SCL
A1.Pd Pr
SPr.2
A1.OP LAt
dPnG
Nl
A2.HL Hl
A2.Pd Pr
A2.OP OFF
Unit
F
Changing the Controller’s Parameter Settings
1. Press the key to enter the programming mode. The lower display will
alternately display the menu level and “Ac.Cd.”
2. Use the
and
keys to change to the desired menu level.
3. Once you have chosen the desired menu use the
key to scroll through
the parameters. To change the setting of a given parameter, use the and
keys.
4. To save settings press the key. The controller now exits the programming
menu and return to the normal operating mode.
5. To change settings on other menu levels, you must re-enter the programming
menu (from step #1).