FEATURES

1) Super large LCD simultaneously display of HCHO level,

Temp., Relative humidity, Calendar (Y/M/D) and Time
(clock).

2) Three (3) smiley icons indicate indoor air quality levels

(0/0.08/0.8ppm), easy to  understand HCHO concentration.

3) Stable Electrochemical type sensor for HCHO detection.
4) High alarm threshold is selectable.
5) Manual 0ppm calibration.
6) Audible/Visible LED warning for high HCHO concentration

1

o

ppm

AIR

TEMP

Humidity

M

D

LCD DISPLAY

Symbols:
1)ppm              HCHO unit
2) icon

: 0ppm ~ 0.08ppm

3) icon

: 0.08ppm ~ 0.8ppm

4) icon

: 0.8ppm

5)Air Temp.: Ambient Temperature
6)MD : Month / Date
7)HM: Hour / Minute
8)TWA: Time Weighted Average

(8 hours)

9)STEL: Short-Term Exposure Limit

(15 minutes weighted average )

10)Logging: When it is in the recording mode

11)VOC, NO2, WBGT, hpa, mbar ,EXT TEMP,lps%,cfm/p are not

available in these models

:

INTRODUCTION

Thank you for purchasing desktop HCHO meter.

It is used to measure HCHO concentration,air temperature

and relative humidity with visible and audible alarms.

This HCHO meter is an ideal instrument for indoor air quality

(IAQ) diagnosis and HVAC system performance verification.

MATERIAL SUPPLIED

(1) Meter

(2) Universal Adaptor

(3) Operation manual

(4) 4 pcs batteries.

(5) USB cable

(6) White Box

POWER SUPPLY

The meter is powered by Super heavy duty batteries AAA x 4 pcs

(Recommended) or an AC adaptor(9V/0.5 A output)

2

KEYPAD (CONTROLS)

1) MODE    : HCHO mode,MAX,TWA,STEL

2) SEL./R   : Edit alarm function

Logger mode: Long press to key start/off.

3) TRH.M.  : Temp. Humidity mode. Setting/ Next

4) UNIT

: Select Temp. unit

5) POWER: Turn meter on/off, alarm on/off.

OPERATION

Warning: Remove batteries when not in use to prevent battery

leakage (Out of warranty) and power consumption.

(a) POWER ON/OFF

Put battery or plug in adaptor, meter turns on automatically .

LCD shows current HCHO, RH & Temperature, Date and Time.

Three(3) smiley icons indicate the indoor air quality level and

appear on the top of first layer display.

Note:

(1) When low battery, meter will restart and logger will stop if

user switch to adaptor.

(2) Please use properly power  source, input voltage:

100~240   VAC,50-60Hz

Output voltage: DC 7.5~9.0V

Output current: 0.5A,or the meter will be damaged.

(3) Battery life approx. 120 hours when continuous used

(b) TAKING MEASUREMENT

The meter starts the measurements when powered

on and response time is less than 30 sec.

Do not hold the meter close to any other source of HCHO.

Note:

(c)  HCHO MODE

MODE

In the HCHO Mode, press

button to switch HCHO

display. The top layer of display will cycle from HCHO,

Maximum, TWA, STEL.

3

Note 1:

1) MAXIMUM:

The unit automatically records maximum reading

and minimum reading after powered on.(MINIMUM is

always 0.00ppm for HCHO meter)

2) TWA: ( Time Weighted Average 8 hours)

The meter keeps updating the reading every minute. If

the meter has been powered on for less than 8 hours,

the TWA value will be the weighted average of readings

since powered on.

3)STEL: (Short-Term Exposure Limit in 15 minutes weighted

average)The meter keeps updating readings every

minute. If the meter has been powered on for less than

15 minutes,  the STEL value shows the weighted average

of reading taken since powered on.

4)RESET MAX:

Press

button to get MAX reading and

MODE

switch TWA or STEL.Under MAX mode, Long

press

button  to clear all the values.

MODE

ppm

MAX

ppm

TWA

STEL

ppm

ppm

(d) TRH MODE : Temperature / Relative Humidity mode

TRH

In the

Mode, the meter displays both

related humidity and air Temperature

simultaneously.

(e) Temperature unit (°C or °F) selection

Press

button to toggle the

UNIT

temperature unit (°C or°F).

o

AIR

TEMP

HUMIDITY

+ TEMPERATURE

HUMIDITY

o

AIR

TEMP

HUMIDITY

o

AIR

TEMP

HUMIDITY

4

MAINTENANCE

Cleaning and storage:

1). The meter should be cleaned with a damp cloth and mild

detergent when necessary.

2). Store the meter in an area with moderate temperature

and humidity.

MANUAL CALIBRATION TO ZERO

Before start HCHO manual calibration make sure the area is

ventilation and no HCHO environment.  Ensure the calibration

completed, to use adaptor during calibration is recommended.

HCHO meter is completed calibration in factory before

packed. Manual calibration is only for meter inaccuracy.

When power on, long press

button more

UNIT

than 6 seconds, the value will turn into “00000"

and flash

seconds.

60

After calibration completed, display will show

full LCD icon for a second. Meter will restart

automatically to return normal measurement.

BATTERY INDICATION

When battery is low (HCHO consumes power ),the reading

of HCHO will be terribly high, could be couple thousand in

ppm. “LO BAT” is flashing at the bottom of the screen,

near calendar and time. Replace new battery when LO BAT

shows.

SETTING DATE & TIME

a) Real time and date:

The meter shows Mo./Date and Hr/Min on the 3rd layer of

display, and each cycle is 16 seconds.

b) Month/Date and Hour/Min

Press

button to enter the real date and

MODE+POWER

time setting. The time default is 24-hours format.

M

D

5

HCHO Hi alarm setting.

Enter HCHO alarm setting.

+

SEL/R POWER

- -> 

- ->

 

- ->

 

- ->

 

Change digit.

UNIT

Number goes up.

SEL/R

Number goes down.

TRH/M

After set up

+

to save and exit setting.

SEL/R POWER

ALARM SETTING

Set up alarm function:

In normal measurement, press

buttons ,

SEL/R+POWER

bell will appear alarm function. Once set up, the red

LED will flash if the data exceeding the setting data.

SETTING DATE & TIME

The meter interchangely shows current date and time

on the 3rd layer of display, and each cycle is

6 seconds.

The time default is 24-hours format.

+

-->Enter date/time mode

MODE POWER

--> Number goes up.

SEL/R

--> Number goes down.

TRH/M

-->Edit Month/date or Hour/Minute.

MODE

-->Select to edit Month or date, Hour or Minute.

UNIT

After set up, press

+

to save and exit.

MODE POWER

HCHO alarm

on

HCHO alarm

off

o

ppm

AIR

TEMP

Humidity

M

D

o

ppm

AIR

TEMP

Humidity

M

D

HCHO  alarm ON/OFF:

In normal function, short press

bell icon appear

POWER

alarm on, short press

again to turn off.

POWER

o

AIR

TEMP

Humidity

H          M

TEMP. HUMIDITY OFFSET

Make sure the meter is off, press

button

SEL/R+TRH/M

then plug in the adaptor at the same time to enter offsetting

mode.

WARNING: DO NOT USE THIS FUNCTION WITHOUT

STANDARD INSTRUMENT.

Example: Standard instrument is 30.2°C,

meter shows 30.0°C.Enter temp. offset mode, press

till 0.2 appear.

TRH/M

Offset range of Temp.and RH as below:

Temp.

:-9.9°C to +9.9°C  /  -9.9°F to +9.9°F

RH%

:-9.9% to +9.9%

a) Humidity offset:

Increase value

Decrease value

b) Air temp. offset

Increase value

Decrease value

After setting up   short press

to save setting and

,

POWER

“%”

icon will flash a second. Then replug adaptor to return normal

measurement.

Logger setup:

When using datalogger function, please use batteries and adaptor

at the same time to avoid power outages and data loss.

A. Install logger driver for Windows

The PL2303_Prolific_DriverInstaller_v110.exedriver is for

Windows operation system, follow the installation steps

hereunder:

1.Put the CD into CD driver, click

PL2303_Prolific_DriverInstaller_v110.exe for installation the

driver.

2.The set up status.

3.Click “NEXT” to continue.

4.After complete the installation, press “Finished”.

6

7

B. Install the logger software for Windows XP

(win7/win8/VISTA)

Please follow the following steps to install the software

1. Put CD into CD driver, click “Setup.exe” to install

2.Select the destination folder you are go to save by click

Browse button

3.Once you finish selecting ,press Next

”.

4.Select “I accept the License Agreemen:(s)” and press NEXT

button.

5.Click NEXT button when you see the following.

6.Press FINISH button to complete the installation.

7.Restart the Windows XP system

C Run the software from Windows XP

.

1. Plug the USB cable into USB port, since you have already

Installed the USB driver, computer will automatically detect

logger ,

2. Find “Datalogger” in START      PROGRAMS

3.Select logger software folder.

Reminder:

Software is installed completed,”Get Datalog’s Identifier”shows

next page(pic.1), then press SETTING button to enter Log setting

(Pic.1)

Note: If logger is not plugged in USB port, the warning message

“Reminder: Data Logger is not plugged in USB port!” shows next

page(pic.2) Make sure logger is plugged in and choose the

correct COM port.

MADE IN TAIWAN

AQM-101

IAQ Formaldehyde 

(HCHO) Logger 

M5443/1114

INSTRUCTION

SHEET

Shop online at: www.omega.com   e-mail: [email protected]