*ACCURACY:
Includes repeatability, temperature
coefficient (18° to 28° C), time stability
(1 year), and linearization conformity
errors. Excludes probe errors; however,
probe errors around 0° C may be compen-
sated by an internal adjustment.
REPEATABILITY:
0.1° C typical for 1 week at constant ambi-
ent temperature.
TEMPERATURE COEFFICIENT:
18° to 28° C; included in accuracy specifica-
tion. From –10° to 18° and 28 ° to 50° C:
less than
6
(0.06% rdg + 0.01° C)/° C.
SENSOR VOLTAGE:
40 mV RMS maximum.
MODEL 865 AND 866 GENERAL SPECIFICATIONS (unless specified):
DISPLAY:
3
1
⁄
2
-digit LCD O.5" (13 mm) height. Polarity
and decimal point indication.
CONVERSION RATE:
1.5 readings per second.
OPEN SENSOR INDICATION:
Display reads between –100° and –199°.
OVERRANGE INDICATION:
3 least significant digits blanked.
MAXIMUM COMMON
MODE VOLTAGE:
42 V peak to earth.
COMMON MODE REJECTION
(Model 865):
Less than 0.01°F/volt at dc, 50 and 60 Hz.
(2k
V
unbalance, LO driven)
COMMON MODE REJECTION
(Model 866):
Less than 0.01°C/volt at dc, 50 and 60 Hz.
(2k
V
unbalance, LO driven)
ENVIRONMENTAL LIMITS FOR
OPERATING (Model 865):
14° to 122°F, less than 80% relative
humidity up to 95°F; linearly derate 1.5%
R.H./°F from 95° to 122°F.
ENVIRONMENTAL LIMITS FOR
OPERATING (Model 866):
–10° to 50°C, less than 80% relative
humidity up to 35°C; linearly derate
3% R.H./°C from 35° to 50°C.