C-3
Appendix - Determining an Unknown Emissivity
C
Method 4
1.
Paint a sample of the object material with flat black lacquer paint.
2.
Set the emissivity to 0.97 and measure and record the temperature
of the painted portion of the sample material - Area ‘A’ in Figure
C-1. Make sure that the painted area of object material fills the
FOV of the thermometer.
3.
Aim the thermometer at another spot on the target - Area ‘B’ in
Figure C-1.
4.
Adjust the emissivity of the thermometer until the temperature
reading equals the temperature found in Step 2.
Method 5
1.
Use this method where practical to measure objects at
temperatures above 260°C (500°F ).
2. Drill a 35 mm (1.5") diameter hole in a sample of the object
material to a depth of 127 mm (5"). This hole closely resembles a
blackbody (refer to Appendix A).
Figure C-2. Determining Emissivity with a Drilled Hole
3.
Set the emissivity to 0.97 and measure and record the temperature
of the hole in the sample material - Area ‘A’ in Figure C-2. Make
sure that the hole fills the FOV of the thermometer.
4.
Aim the thermometer at another spot on the target as close as
possible to Area ‘A’ (Area ‘B’ in Figure C-2).
5. Adjust the emissivity of the thermometer until the temperature
reading equals the temperature found in Step 3.
Drilled out
Area 'A' Target
Area 'B' Target
Summary of Contents for OMEGASCOPE OS523E
Page 3: ...Unpacking Instructions n4 Notes...
Page 5: ...ii...
Page 15: ...General Description 1 1 8 Notes...
Page 61: ...7 4 Troubleshooting Guide 7 Notes...
Page 67: ...Specifications 8 8 6 Notes...
Page 69: ...Glossary of Key Strokes 9 9 2 Notes...
Page 79: ...Appendix Determining an Unknown Emissivity C C 4 Notes...