6
Maintenance
6-1
6.1 Introduction
•
No lubrication is required for the pump. All bearings
are pre-sealed and rated for long life.
•
After many hours of use, fine particles of tubing will
tend to accumulate inside the pump and on the rollers.
Use a high pressure air hose (60 PSI) to blow out most of
the particles from the pump. Clean all parts with a mild
soap solution or a light mineral oil.
•
The pump may be dismantled either for cleaning or for
replacing the rotor assembly in case of a malfunction.
Follow the procedure in Section 6.2 to replace the rotor
assembly.