7-8
NetScan User’s Manual
Configuration Settings on Rear Panel DIP Switch
Micro-
Label
Setting
Description
switch #
1
COMM SELECT
0
Selects serial communication
1
Not applicable to NetScan
2,3
HANDSHAKE (H/S)
1
00
No Handshake
01
Software Handshake only
2
(XON/XOFF)
10
Hardware Handshake only
(RTS/CTS)
11
Both Hardware/ Software
Handshake
4,5
PARITY
00
No Parity
01
Odd Parity
10
Even Parity
11
(N/A)
6,7,8
SERIAL BAUD RATE
000
300 baud
001
600 baud
010
1200 baud
011
2400 baud
100
4800 baud
101
9600 baud
1
110
19200 baud
1
111
(N/A)
9
RS-232
0
Selects RS-232 Serial
RS-422
1
Selects RS-422 Serial
Note 1: At 9600 and 19200 baud rates, Handshake should be set for “Hardware Handshake
only,” 10.
Note 2: ChartView software will not run when Handshake is set for “Software Handshake
only,” 01.
Serial Port Pin Connectors
NetScan is equipped with two DB9 connectors, both located on the rear panel. The DB9
connector labeled CONFIG is for ethernet configuration only. The DB9 connector labeled
RS-232/422 is for serial interface applications. The serial port default is for the RS-232
connector configured as DTE (Data Terminal Equipment) for an IBM AT at RS-232 levels.
An optional CA-47 cable (or equivalent) connects the NetScan with the computer. The NetScan side of the cable has
a male DB9 connector for use with the RS-232/422 serial port. The cable’s computer side has 2 connectors: one for
DB9, and the other for DB25. Other crossover-type cables can be used if wired as shown in the figure. The
following table lists the RS-232 and RS-422 pin connections for a DB9.
Pin #
RS-232
Pin Description
Signal
1
No connection
N/A
2
Receive Data (RxD-)
Input
3
Transmit Data (TxD-)
Output
4
No connection
N/A
5
Ground
Ground
6
No connection
N/A
7
Request to send (RTS-)
Output
8
Clear to send (CTS-)
Input
D B 9
2
7 8
3
5
D B 2 5
2
7
3
5
4
C o m p uter N etS c an
D B 9
R X D 2
TX D 3
G N D 5
RT S 7
C TS 8
3 TX D
2 R X D
5 G N D
8 C T S
7 R T S
D B 9
D B 2 5
C o m p uter N etS ca n
TX D 2
R X D 3
G N D 7
C TS 5
RT S 4
3 TX D
2 R X D
5 G N D
8 C T S
7 R T S
D B 9
Interface W iring Diagram
9
No connection
N/A
Summary of Contents for OMB-NETSCAN 1500
Page 6: ...iv NetScan User s Manual...
Page 18: ...1 12 Configuring and Starting NetScan NetScan User s Manual Notes...
Page 38: ...3 8 General Information and Specifications NetScan User s Manual Notes...
Page 82: ...4 44 ChartView Software Reference NetScan User s Manual Notes...
Page 118: ...6 20 Calibration NetScan User s Manual...
Page 140: ...A ii NetScan User s Manual...
Page 192: ...API Command Reference Appendix A A 52 NetScan User s Manual Notes...
Page 237: ...Appendix D Registers Data Formats Queries NetScan User s Manual D 13...
Page 244: ...NetScan Program Examples Appendix E E 2 NetScan User s Manual...
Page 248: ...ASCII Code Summary Appendix F F 4 NetScan User s Manual Notes...
Page 250: ...NetScan Error Messages Appendix G G 2 NetScan User s Manual Notes...
Page 252: ...Abbreviations Appendix H H 2 NetScan User s Manual Notes...
Page 254: ...NetScan User s Manual...