LVR300 Series Quick Reference – Programming Steps

+ 0.0

Inch (cm)

S1

+25.5

S1

S1

S1

S1

S1

S1

+18.6

+18.6

+28.6

+21.6

+25.6

+25.

S1

Min.

Cursor flashing

2x

3x

9x

HOW TO MODIFY THE SENSOR

OPERATING PARAMETERS

Example: S1 = +18.6 Inches (or cm)

HOW TO ENTER ADDITIONAL

PARAMETERS & INFORMATION

+ 0.0

Inch (cm)

Code

0 0 1

Code

Code

Filter

Units

4 mA

0

Inch (cm)

1

0 1 1 

0 1 1 

Output

4-20 mA

20 mA

80

Turning left
moves the cursor
to the next deci-
mal position.

By turning the ring
to the left toward
Position 1, 
the cursor disap-
pears and enters
the code.  The
next parameter
arrives automati-
cally.

Code

0 0 0

Code

0 0 1

Code

1 1 1

T

O

 R

ETURN TO THE

 F

ACTORY

 D

EFAULT

 S

ETTING

:

Set the code to 989.

All individual parameters are now overwritten.

Start of analog 
output range 
(in the selected unit).

Set analog output 
4-20 mA or 0-20 mA.

Selection of various
units (inches, cm, PSI).
Display and switch
points are recalculated
automatically.

End of analog 
output range 
(in the selected unit).

Filter for display and
analog output. (It is a
FIFO filter and the
time to accomplish the
actual measurement
value corresponds to 
the seconds shown in
the display.)

In regular display
mode again.

Turning right
effects an increase
of the figure.

If the ring is turned
right toward Position 2
(EDIT), a cursor
appears while the
parameter you want to
modify is displayed.

Turn the ring 2 times to
the left, toward
Position 1, and the cur-
sor disappears.  The
modified S1 is now vis-
able and the next para-
meter is shown
automatically.

Move ring 3 times to
the right, alternating to
Position 2 and center,
and you will see a 
3-digit increase to the
figure marked by the
cursor’s location.

Turning left moves the
cursor to the next deci-
mal position.

Turning right effects 
a change to the next
higher figure.

When the program
ring is turned to
the right toward
Position 2 (EDIT),
a cursor appears
at the parameter
to modify.

NOTE: If you remain idle in the programming stage
for 30 seconds, the display indication will return to
“measurement display” without saving your new set-
tings. You will have to cycle through the program
again to make your changes.

Summary of Contents for LVR300 Series

Page 1: ...omega com e mail info omega com For latest product manuals omegamanual info LVR300 Series Continuous Level Switch System Shop online at User sGuide LVR300 Series...

Page 2: ...78280 Guyancourt France TEL 33 0 1 61 37 2900 FAX 33 0 1 30 57 5427 Toll Free in France 0800 466 342 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany TEL 49 0 7056...

Page 3: ...nt accuracy The signal is displayed with units using a back lit LCD graphical display and converted to a 4 20 mA signal Two switch points with either a positive or negative output can be programmed ov...

Page 4: ...s threshold added to switch point 1 for switch active Hysteresis threshold added to switch point 2 for switch active Returns to regular display of current float level To access other program para mete...

Page 5: ...output range in the selected unit Filter for display and analog output It is a FIFO filter and the time to accomplish the actual measurement value corresponds to the seconds shown in the display In r...

Page 6: ...2 The pressures and temperatures must take into consideration possible surges in the temperature and pressure of the system 3 The liquids used must be compatible with the materials of construction Sp...

Page 7: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Page 8: ...ity Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial...

Reviews: