The Orbit Network has been designed to meet EMC requirements EN50081-1 for emissions and 

EN50082-1 for immunity when properly installed. The T-CON case should be connected to a good elec

-

trical ground. The mounting option using the captive nuts is the preferred method. Ensure screws make 

good contact with the mounting surface. If the mounting option using plastic feet is used paint should be 

removed from the T-CON base pads to ensure good electrical contact with mounting surfaces.

Cable

The Gage Head cable is custom made to achieve the optimum balance between flexibility, flex life, 

chemical resistance, abrasion resistance electrical and electrical screening.

To minimise transducer failure due to cable damage cable runs should be positioned well clear of moving 

components and vulnerable working areas. If the cable is in a flex situation then a minimum bend radius 

of 150 mm should be maintained, the vacuum and pneumatic options should not be flexed from the 

transducer cable entry, but should be anchored separately at a position of 50 - 70 mm from the end cap.

If the cable of a Digital probe is damaged it is not possible to repair it without affecting probe calibration. It 

must be replaced.

Connections

The PIE is designed to be used with the T-CON connector and allows for fast network configuration. All 

connections are via the 9-way sub D-type connector.

Summary of Contents for GP901 Series

Page 1: ...Series 956 User s Guide http www omega com e mail info omega com GP901 Series Digital Probes...

Page 2: ...69 6311114 Toll Free 0800 1 66342 e mail czech omega com France 9 rue Denis Papin 78190 Trappes Tel 33 130 621 400 FAX 33 130 699 120 Toll Free in France 0800 4 06342 e mail france omega com Germany A...

Page 3: ...first are the specifications for the complete Digital Probe PIE gauge head and the second is mechanical data relevant to the gauge head only PIE is designed to be primarily used with the T CON connect...

Page 4: ...S 6 0 05 mm 6 0 2 x D 1 mm Probes i e GP901 1 S 6 0 1 mm 6 0 1 x D 2 mm Probes i e GP901 2 S GP901 2 P 6 0 1 mm 6 0 1 x D 5 mm Probes i e GP901 5 S GP901 5 P 6 0 1 mm 6 0 2 x D 10 mm Probes i e GP901...

Page 5: ...ward direction and the extension of the tip is achieved by applying air pressure to the Gauge Head nozzle Air pressure may be varied within the specified limits to achieve the desired tip force at zer...

Page 6: ...lity flex life chemical resistance abrasion resistance electrical and electrical screening To minimise transducer failure due to cable damage cable runs should be positioned well clear of moving compo...

Page 7: ...GP901 M 3343 05 08 PIE and T CON Installation...

Page 8: ...GP901 M 3343 05 08 This Page Intentionally Left Blank...

Page 9: ...GP901 M 3343 05 08 This Page Intentionally Left Blank...

Page 10: ...10 GP901 M 3343 05 08 This Page Intentionally Left Blank...

Page 11: ...otherwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sol...

Page 12: ...cators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industrial pH Cond...

Reviews: