background image

        D M D - 4 8 0 - S 2   S E R I E S     A C   P O W E R E D   B R I D G E S E N S O R   -   R S 2 3 2   I N T E R F A C E

6

PHONE:(203)359-1660  FAX:(203)359-7700

1-800-TC OMEGA

[email protected]

6.

OPTION 5 [SELECT UNITS TO MONITOR]:

THE LAST OPTION SELECTS THE ADDRESS OF THE MODULES TO
BE MONITORED IN THE ORDER IN WHICH THEY ARE TO BE MONITORED.
ADDRESSES FROM 0 TO 63 MAY BE SELECTED.

THE DEFAULT ADDRESS IS “1”.

E.

OPERATION  MENU

THE “OPERATION MENU”, ALLOWS 4 USER CHOICES.

1.

OPTION 0 [QUIT]:

ALLOWS THE USER TO QUIT THE PROGRAM.

2.

OPTION 1 [SETUP THE UNITS]:

EXECUTES THE “SETUP MENU”, WHICH ALLOWS THE USER TO
SET ADDRESSES TO INTERROGATE, MEASUREMENT RATE, ETC.

REGARDLESS OF WHETHER THE “TEST MODE” OR “OPERATION

MODE”

IS SELECTED, CERTAIN PARAMETERS MUST ALWAYS BE ENTERED IN
THE SETUP UNITS OPTION IN ORDER FOR YOUR MEASUREMENT
SYSTEM TO OPERATE PROPERLY.

(REFER TO SECTION D FOR SETUP MENU DESCRIPTION)

3.

OPTION 2 [START OPERATE MODE]:

WILL EXECUTE THE OPERATE MODE IF THE DMD-480-S2 SERIES HAS
BEEN SET FOR MODE 0 (SW7 AND SW8 “ON”).

4.

OPTION 3 [START TEST MODE]:

WILL EXECUTE THE TEST MODE IF THE DMD-480-S2 SERIES HAS BEEN
SET FOR MODE 1 OR 3.

PRIOR TO ENTERING THE OPERATE MODE, IT WILL BE NECESSARY TO HAVE THE
CORRECT UNIT ADDRESSED, AND IN MOST CASES IT WILL BE NECESSARY TO
ENTER THE SETUP MENU AND SET THE SCALING EQUATION, DATA
DESTINATION, AND DATA COLLECTION METHOD.

IT IS RECOMMENDED THAT THE USER OPERATE THE UNIT IN TEST MODE 3 PRIOR
TO THE OPERATE MODE TO ENSURE THAT THE UNIT IS COMMUNICATING
PROPERLY.  THE USER CANNOT ADDRESS MULTIPLE UNITS IN THE TEST MODE,
THEREFORE EACH UNIT MUST BE TESTED INDIVIDUALLY.

F.

SETTING  THE  SCALING  EQUATION

THE SCALING EQUATION IS SET BY WRITING THE EQUATION FOR THE TWO
CONDITIONS OF ZERO AND SPAN, AND THEN SOLVING FOR THE CONSTANTS.
THE DMD-480-S2 HAS AN OUTPUT OF APPROXIMATELY 41,750 COUNTS FULL
SCALE.  IF THE MODULE WERE TO BE USED AS A VOLTAGE OUTPUT UNIT AND
THE SPAN WERE TO BE 30 mV, THEN THE FOLLOWING CALCULATION WOULD BE
MADE.

Summary of Contents for DMD-480-S2 Series

Page 1: ...4500 BAS PROGRAM USER S MANUAL PHONE 203 359 1600 1 800 TCOMEGA FAX 203 359 7700 info omega com DMD 480 S2Series AC POWERED BRIDGESENSOR RS232 INTERFACE V e r s i o n 2 1...

Page 2: ...p 1 e t a R n o i s r e v n o C d n o c e s s e m i t 5 t u p n I o t d e r r e f e R y t i r a e n i L e l a c S l l u F f o 1 0 0 C 5 5 o t 0 t n e i c i f f e o C e r u t a r e p m e T l a c i p y...

Page 3: ...OMPUTERS SERIAL PORT COM1 OR COM2 ONLY AND THE OTHER END TO THE DMD 480 S2 SERIES FEMALE DB9 CONNECTOR THE RECOMMENDED MAXIMUM CABLE LENGTH IS 50 FT WHICH IS DETERMINED BY THE MAXIMUM SPECIFIED CABLE...

Page 4: ...E OMEGA C4500 SOFTWARE QBASIC OR GWBASIC MUST BE LOADED ON YOUR COMPUTER 2 TO BEGIN THE SOFTWARE SETUP YOU MUST LOAD AND EXECUTE THE SUPPLIED OMEGA PROGRAM C4500 BAS WITHIN QBASIC OR GWBASIC D SETUP M...

Page 5: ...N THE SCALING EQUATION PROVIDES FOR BOTH A SCALE FACTOR A AND A FIXED OFFSET C THE SCALING EQUATION IS VALUE C A x READING THE SETTING OF THE SCALING EQUATION ONLY APPLIES TO THE OPERATION MODE REFER...

Page 6: ...ION 2 START OPERATE MODE WILL EXECUTE THE OPERATE MODE IF THE DMD 480 S2 SERIES HAS BEEN SET FOR MODE 0 SW7 AND SW8 ON 4 OPTION 3 START TEST MODE WILL EXECUTE THE TEST MODE IF THE DMD 480 S2 SERIES HA...

Page 7: ...5 COUNTS WITH 30 mV APPLIED THE DMD 483 S2 OUTPUT READS 41 745 COUNTS A 41 745 x A C 0 030 VOLTS B 4185 x A C 0 003 VOLTS SOLVING EQUATION A EQUATION B 37 560 x A 0 027 VOLTS THEREFORE A 0 027 37 560...

Page 8: ...ATH DRIVE PATH FILE THE DATA RECORDED IN THE FILE IS RECORDED IN THE FOLLOWING MANNER UNIT NUMBER ADDRESS DATE TIME UNSCALED READING AND THE SCALED READING AN EXAMPLE OF A FILE S CONTENTS IS SHOWN BEL...

Reviews: