Omega Engineering DFG35 Series User Manual Download Page 4

DFG35 Digital Force Gauges 

 

User’s Guide 

 
 

 

 

 

3

 

1   OVERVIEW  

 

1.1  List of included items 

 

Qty. 

Part No. 

Description 

DFG35-0.12 – 

DFG35-20 

DFG35-50 – 

DFG35-100 

DFG35-200 – 

DFG35-500 

12-1049 12-1049 12-1049 Carrying Case 

08-1022 08-1022 08-1022 AC adapter body with US, EU, or UK prong 

08-1026 08-1026 08-1026 Battery (inside the gauge) 

G1024 G1024 G1031 Extension 

rod 

G1026 G1026 G1033 Cone 

G1025 G1025 G1032 Chisel 

G1027 G1027 G1034 V-groove 

G1029 G1029 G1036 Flat 

G1028 G1038 G1035 Hook 

1 N/A G1039 

G1037 

Coupling 

Certificate of calibration 

1 09-1165 

USB 

cable 

1 - 

Resource 

CD 

(USB driver, user’s guide) 

 

1.2 Safety / Proper Usage 

 

Caution! 

Note the force gauge’s capacity before use and ensure that the capacity is not exceeded. 
Producing a force greater than 150% of the gauge’s capacity can damage the internal load cell. An 
overload can occur whether the gauge is powered on or off. 

 
Typical materials able to be tested include many manufactured items, such as springs, electronic 
components, fasteners, caps, films, mechanical assemblies, and many others. Items that should not be 
used with the gauge include potentially flammable substances or products, items that can shatter in an 
unsafe manner, and any other components that can present an exceedingly hazardous situation when 
acted upon by a force. 
 
The following safety checks and procedures should be performed before and during operation: 
 

1.  Never operate the gauge if there is any visible damage to the AC adapter or the gauge itself.  

 

2.  Ensure that the gauge is kept away from water or any other electrically conductive liquids at all 

times. 

 

3.  The gauge should be serviced by a trained technician only. AC power must be disconnected and 

the gauge must be powered off before the housing is opened. 

 

4.  Always consider the characteristics of the sample being tested before initiating a test. A risk 

assessment should be carried out beforehand to ensure that all safety measures have been 
addressed and implemented. 

 

5.  Wear eye and face protection when testing, especially when testing brittle samples that have the 

potential to shatter under force. Be aware of the dangers posed by potential energy that can 
accumulate in the sample during testing. Extra bodily protection should be worn if a destructive 
failure of a test sample is possible. 
 

Summary of Contents for DFG35 Series

Page 1: ...DFG35 Digital Force Gauges...

Page 2: ...DFG35 Digital Force Gauges User s Guide 1...

Page 3: ...years of service in laboratory and industrial environments This User s Guide provides setup safety and operation instructions Dimensions and specifications are also provided For additional information...

Page 4: ...thers Items that should not be used with the gauge include potentially flammable substances or products items that can shatter in an unsafe manner and any other components that can present an exceedin...

Page 5: ...he AC adapter is plugged in an icon appears in the lower left corner of the display as follows If the AC adapter is not plugged in battery power drainage is denoted in a five step process 1 When batte...

Page 6: ...e back of the housing is provided to withstand the load during a test A mating dowel pin should be used see illustration below There are four threaded holes located near the corners of the housing The...

Page 7: ...nloaded from www mark 10 com Caution Install the USB driver before physically connecting the gauge to a PC with the USB cable Further instructions for configuring and using the gauge s outputs are pro...

Page 8: ...r conditions Refer to the Power section for details 8 High low limit indicators Correspond to the programmed set points Indicator definitions are as follows the displayed value is greater than the upp...

Page 9: ...tem without saving Press ESCAPE to revert one step back in the menu hierarchy until back into normal operating mode Refer to the following sections for details about setting up particular functions an...

Page 10: ...the peak capture rate of the instrument Displayed Reading Applies to the primary reading on the display Available settings 1 2 4 8 16 32 64 128 256 512 1024 It is recommended to keep the current readi...

Page 11: ...ide of the housing as shown in the illustration in the Power section Communication is possible only when the gauge is in the main operating screen i e not in a menu or configuration area The current r...

Page 12: ...s Numeric Only selection Individual data points may be transmitted by pressing DATA 10 CALIBRATION 10 1 Initial Physical Setup The gauge should be mounted vertically to a test stand or fixture rugged...

Page 13: ...press ESCAPE The display will appear as follows Selecting CANCEL will revert back to the Calibration setup Selecting EXIT W O SAVING will return to the menu without saving changes 4 After the number o...

Page 14: ...le of the instrument and then press ENTER 9 After displaying PLEASE WAIT the display will appear as follows Remove the load applied in Step 8 leave the fixtures in place then press ZERO 10 The display...

Page 15: ...ows Press ENTER 12 The display will appear as follows Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees Press DIRECTION to invert the display Then attach weight fixtures...

Page 16: ...ing in any manner Then try again The calibration weight does not match the set value The entered calibration point is too close to the previous point CALIBRATION TENSION Load too close to previous Ple...

Page 17: ...cess these settings select Backlight from the menu The display will appear as follows Select Off for the backlight to be off upon powering on the gauge Select On for the backlight to be on upon poweri...

Page 18: ...ply select Tones from the menu The screen will appear as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the gauge To access this parameter select Initial...

Page 19: ...7 lb 0 33 kg DFG35 200 DFG35 500 0 9 lb 0 41 kg Included accessories Carrying case chisel cone V groove hook flat extension rod AC adapter battery USB cable resource CD USB driver and user s guide NI...

Page 20: ...0 12 50 0 5 0 0001 0 05 0 0005 0 005 0 13 DFG35 0 25 0 25 100 1 0 0002 0 1 0 001 0 010 0 25 DFG35 0 5 0 5 250 2 5 0 0005 0 2 0 002 0 010 0 25 DFG35 2 2 1 10 0 002 0 001 0 01 0 010 0 25 DFG35 5 5 2 5 2...

Page 21: ...DFG35 Digital Force Gauges User s Guide 20...

Page 22: ...DFG35 Digital Force Gauges User s Guide 21 M 5085 1017...

Reviews: