Omega Engineering DFG35 Series User Manual Download Page 15

DFG35 Digital Force Gauges 

 

User’s Guide 

 
 

 

 

 

14

 

 

Use the 

UP

 and 

DOWN

 keys to adjust the load value as required. The load values default to even 

increments, as indicated by the previously entered number of data points (even increments are 
recommended for best results). For example, if a 50 lbF capacity gauge is calibrated, and 5 data 
points were selected, the load values will default to 10, 20, 30, 40, and 50 lb.  Apply the 
calibration load. Then press 

ENTER

 

Repeat the above step for the number of data points selected.  

 

11.  After all the compression calibration points have been completed, the display will appear as 

follows: 

          

 

Press 

ENTER

 

12.  The display will appear as follows: 

 

 

Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees. Press 

DIRECTION

 to invert the display. Then attach weight fixtures. The following screens will step 

through the same procedure as with the compression direction. Proceed in the same manner.  
 

13.  At the completion of the tension calibration, the display will appear as follows: 

           

 

To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select 
“EXIT W/O SAVING”. 

 

14.  Any errors are reported by the following screens:  

 

CALIBRATION 

 

   To invert the 
   display, press the 
   DIRECTION button, 
   then press ENTER 
 

CALIBRATION 

COMPLETE 

 
   

Save & exit

 

   Exit w/o saving 
 

CALIBRATION 

COMPRESSION COMPLETE 

       Reverse direction  
       for tension. 
       Attach necessary 
       weight fixtures, 
       then press ENTER. 

Summary of Contents for DFG35 Series

Page 1: ...DFG35 Digital Force Gauges...

Page 2: ...DFG35 Digital Force Gauges User s Guide 1...

Page 3: ...years of service in laboratory and industrial environments This User s Guide provides setup safety and operation instructions Dimensions and specifications are also provided For additional information...

Page 4: ...thers Items that should not be used with the gauge include potentially flammable substances or products items that can shatter in an unsafe manner and any other components that can present an exceedin...

Page 5: ...he AC adapter is plugged in an icon appears in the lower left corner of the display as follows If the AC adapter is not plugged in battery power drainage is denoted in a five step process 1 When batte...

Page 6: ...e back of the housing is provided to withstand the load during a test A mating dowel pin should be used see illustration below There are four threaded holes located near the corners of the housing The...

Page 7: ...nloaded from www mark 10 com Caution Install the USB driver before physically connecting the gauge to a PC with the USB cable Further instructions for configuring and using the gauge s outputs are pro...

Page 8: ...r conditions Refer to the Power section for details 8 High low limit indicators Correspond to the programmed set points Indicator definitions are as follows the displayed value is greater than the upp...

Page 9: ...tem without saving Press ESCAPE to revert one step back in the menu hierarchy until back into normal operating mode Refer to the following sections for details about setting up particular functions an...

Page 10: ...the peak capture rate of the instrument Displayed Reading Applies to the primary reading on the display Available settings 1 2 4 8 16 32 64 128 256 512 1024 It is recommended to keep the current readi...

Page 11: ...ide of the housing as shown in the illustration in the Power section Communication is possible only when the gauge is in the main operating screen i e not in a menu or configuration area The current r...

Page 12: ...s Numeric Only selection Individual data points may be transmitted by pressing DATA 10 CALIBRATION 10 1 Initial Physical Setup The gauge should be mounted vertically to a test stand or fixture rugged...

Page 13: ...press ESCAPE The display will appear as follows Selecting CANCEL will revert back to the Calibration setup Selecting EXIT W O SAVING will return to the menu without saving changes 4 After the number o...

Page 14: ...le of the instrument and then press ENTER 9 After displaying PLEASE WAIT the display will appear as follows Remove the load applied in Step 8 leave the fixtures in place then press ZERO 10 The display...

Page 15: ...ows Press ENTER 12 The display will appear as follows Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees Press DIRECTION to invert the display Then attach weight fixtures...

Page 16: ...ing in any manner Then try again The calibration weight does not match the set value The entered calibration point is too close to the previous point CALIBRATION TENSION Load too close to previous Ple...

Page 17: ...cess these settings select Backlight from the menu The display will appear as follows Select Off for the backlight to be off upon powering on the gauge Select On for the backlight to be on upon poweri...

Page 18: ...ply select Tones from the menu The screen will appear as follows 11 5 Initial Mode This section is used to configure the initial mode upon powering on the gauge To access this parameter select Initial...

Page 19: ...7 lb 0 33 kg DFG35 200 DFG35 500 0 9 lb 0 41 kg Included accessories Carrying case chisel cone V groove hook flat extension rod AC adapter battery USB cable resource CD USB driver and user s guide NI...

Page 20: ...0 12 50 0 5 0 0001 0 05 0 0005 0 005 0 13 DFG35 0 25 0 25 100 1 0 0002 0 1 0 001 0 010 0 25 DFG35 0 5 0 5 250 2 5 0 0005 0 2 0 002 0 010 0 25 DFG35 2 2 1 10 0 002 0 001 0 01 0 010 0 25 DFG35 5 5 2 5 2...

Page 21: ...DFG35 Digital Force Gauges User s Guide 20...

Page 22: ...DFG35 Digital Force Gauges User s Guide 21 M 5085 1017...

Reviews: