Darkfield Condenser Installation and Operation 
Instructions 

7.1  Dry darkfield condenser of A191 

1)  Mounting the dry darkfield condenser 

  Loosen the condenser lock thumb screw on the 

condenser holder and remove the brightfield 
condenser. 

  Install the dry darkfield condenser and tighten 

the condenser lock thumb screw on the 
condenser holder (see 

Fig. 12

)

2)  Centering the dry darkfield condenser 

  Turn the 4X objective to the light path.   

  Turn the condenser focus knob to lower the 

condenser till a dark spot showed in the viewing 
field as shown in 

Fig. 13

 

(a)

  Turn the condenser translational centering 

screws to move the dark spot to the center as 
shown in 

Fig. 13

 

(b)

3)  Place the slide on the stage.   

4)  Raise the condenser all the way to the top and lower it a little bit.   

5)  Following the procedures in this manual to focus and observe. 

6)  Move the condenser up or down slightly to get the best darkfield viewing. 

Note

 

  The dry darkfield condenser is used with the dry objectives only. 

  The dry darkfield condenser works with the 4X, 10X, 40X objectives. 

 

The dry darkfield condenser won

’t work with the 100X oil immersion objective.

 

7.2  Oil darkfield condenser of A191BD, A191BOIL 

1)  Mounting the oil darkfield condensers 

  Loosen the condenser lock thumb screw on the 

condenser holder and remove the brightfield 
condenser. 

  Install the

 

oil darkfield condensers and tighten the 

condenser lock thumb screw on the condenser 
holder (see 

Fig. 14

)

2)  Centering the oil darkfield condensers 

  Turn the 40X objective to the light path.   

  Turn the condenser focus knob to slowly lower 

and raise the condenser till a dark spot showed in 
the viewing field as shown in 

Fig. 15

 

(a)

  Turn the condenser translational centering 

screws to move the dark spot to the center as 
shown in 

Fig. 15 (b)

.

 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

 

     

(a)                                (b) 

 

   

 

   

(a)                                (b) 

Fig.12 

Fig.13 

Fig.14 

Fig.15 

Condenser Lock   

Thumb Screw 

Centering Screws 

Condenser Holder 

Summary of Contents for M827S Series

Page 1: ...User Manual Binocular Biological Compound Microscope Model M827S Series MicroscopeNet com...

Page 2: ...1 Component Illustration 3 2 Installation 4 3 Operation 7 4 Specifications 10 5 Optional Parts 11 6 Troubleshooting Guide 13 7 Darkfield Condenser Installation and Operation Instructions 15 8 Phase C...

Page 3: ...icroscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any other component disconnect the power cord immediately and wi...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...Screw 19 Condenser Lock Thumb Knob 4 Nosepiece 12 Microscope Body 20 Condenser Focus Knob 5 Objective 13 Focus Tension Ring 21 Focus Stop Lever 6 Slide Holder 14 Coarse Focus Knob 22 Abbe Condenser 7...

Page 6: ...mechanical stage is at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the h...

Page 7: ...t You may also loosen the two screws on the cover to adjust the position of the bulb to get centered and even brightness See Fig 3 7 Screw the cover on Caution Before you turn over the microscope be s...

Page 8: ...no power is available 2 8 Installation of the camera optional may not included in your package 1 Take off the eyepiece from the eyepiece tube 2 Insert the camera into the eyepiece tube and then connec...

Page 9: ...pupillary distance While observing with both eyes hold the left and right eyepiece tubes then slide the tubes in and out The interpupillary distance is correct when the left and right fields of view c...

Page 10: ...ctive nosepiece to seat the 100X objective to the observing position until you hear a click sound 4 After observing the specimen use the lens cleaning paper to clean the 100X objective lens gently and...

Page 11: ...ocus tension has been pre set at the factory If the mechanical stage drops by itself rotate the tension adjustment ring Fig 11 situated between the coarse focus knob and microscope body on the power s...

Page 12: ...adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides w focus stop Minimum fine focusing adjustment at 0 002mm range 28mm Mechanical Stage Double layer dimension 5 1 2 x 5 1...

Page 13: ...00X 1 25 oil 0 5 160 0 17 spring 37mm A191BOIL Oil 1 36 1 25 Plan 100X 1 25 oil 0 5 160 0 17 spring w iris diaphragm 37mm 2 Cameras Model Resolution Operating System Software A1502 640 x 480 0 3MP MS...

Page 14: ...ring Achromatic 100X with built in phase plate spring oil five positions 10 for 10X phase contrast objective 20 for 20X phase contrast objective 40 for 40X phase contrast objective 100 for 100X phase...

Page 15: ...too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set proper...

Page 16: ...brightness The condenser is not in the right position Lower or raise the condenser slightly to the position that the specimen is the brightest A191BOIL A191BD only Make sure the oil contact the bottom...

Page 17: ...ondenser up or down slightly to get the best darkfield viewing Note The dry darkfield condenser is used with the dry objectives only The dry darkfield condenser works with the 4X 10X 40X objectives Th...

Page 18: ...jective you will need to drop the immersion oil and adjust the iris ring Fig 16 to get proper brightness and contrast of view field 8 After observing the specimen with 100X objective use the lens clea...

Page 19: ...ighten the thumb screw g Insert the light collector back onto the base and turn until it s tightened h Re adjust the condenser height to its normal position i Re adjust the mechanical stage height to...

Page 20: ...e condenser control knob to raise the condenser to the highest position e Loosen the condenser lock thumb screw Fig 20 pull down the condenser and take it off f Insert the condenser annular ring disk...

Page 21: ...onto the condenser as shown in Fig 22 c Loosen the condenser lock thumb screw as shown in Fig 23 take off the original condenser from the holder d Insert the phase contrast condenser into the condense...

Page 22: ...g is in focus and the dark ring is visible i If the two ring images are not coincided as shown in Fig 24 c adjust the two centering screws on the condenser ring plate till the two rings are coincided...

Reviews: