background image

 

 

www.microscopenet.com

 

 

2

 

Installation   

2.1  Installation of the binocular viewing head 

1)  Loosen the head thumb lock screw on the top of the microscope body and remove 

the plastic cover on the top. 

2)  Remove the cap on the dovetail of the binocular viewing head. 
3)  Seat the dovetail completely of the viewing head into the socket on the top of the 

microscope body and tighten the head thumb lock screw. 

Caution

:   

Do not release the viewing head from your hand grip until you are sure the viewing 
head is installed securely.   

2.2  Installation of the eyepieces   

1)  Remove the protective caps from the eyepiece tubes.   
2)  Insert the eyepieces into the eyepiece tubes.     

2.3  Installation of the objectives 

1)  Adjust the coarse focus knob until the mechanical stage is at its lowest position. 
2)  Turn the caps counter-clockwise to remove them from the nosepiece. 
3)  Take the objectives out from the plastic cases and turn each one clock-wise into the 

holes on the nosepiece. Install the 4X objective into the nosepiece first. Then in a 
counter-clockwise direction, rotate the nosepiece and install each succeeding higher 
magnification objective as shown in 

Fig. 1

   

 
 
 
 
 
 
 
 
 

Note: 

  Inspect the objectives frequently for dirt or oil; clean if necessary. 

  Use the 10X objective to initially focus the image of your specimen.   

  When changing the objective magnification, rotate the objective nosepiece until 

you hear a “click” sound or have a clear "in position" feeling. This ensures the 
objective is centered in the optical light path. 

2.4  Installation of the glass filter 

1)  Swing out the color filter holder under the condenser. 
2)  Place the filter into the holder as shown in 

Fig. 2

, swing the holder in.   

   

 
 

Fig.1 

Summary of Contents for M827L Series

Page 1: ...User Manual Binocular Compound LED Microscope Model M827L Series MicroscopeNet com...

Page 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 2 Components Illustration 3 3 Installation 4 4 Operation 6 5 Specifications 10 6 Optional Parts 11 7 Troubleshooting Guide 12...

Page 3: ...of direct sunlight high temperature or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto...

Page 4: ...ng an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult Oil immersion objectives...

Page 5: ...Screw 19 Focus Stop Lever 4 Nosepiece 12 Microscope Body 20 Condenser Lock Thumb Knob 5 Objective 13 Focus Tension Ring 21 Condenser Focus Knob 6 Slide Holder 14 Coarse Focus Knob 22 Abbe Condenser 7...

Page 6: ...mechanical stage is at its lowest position 2 Turn the caps counter clockwise to remove them from the nosepiece 3 Take the objectives out from the plastic cases and turn each one clock wise into the h...

Page 7: ...take the eyepieces off and be certain that the head is securely locked by the thumb screw 2 6 Installing the mirror optional may not included in your package 1 Turn off the power switch and disconnec...

Page 8: ...on the power switch 3 Rotate the brightness intensity dial to increase or decrease the brightness Caution A diffusion filter is attached beneath the condenser to get uniform light and protect your eye...

Page 9: ...get a good focused image you may need to combine the focus knob adjustment and interpupillary distance adjustment along with eyepiece diopter adjustment stated in 3 3 and 3 4 5 You may now switch to...

Page 10: ...s first 3 7 Adjusting condenser 1 Turn the condenser focus knob to raise or lower the condenser 2 Raise the condenser when using high power objectives and lower it when using low power objectives Note...

Page 11: ...mera optional may not included in your package 1 Bring the microscope into focus by following the procedures in 3 5 2 Install the camera by following the procedures in 2 7 3 Open image observing sofew...

Page 12: ...with iris diaphragm and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on both sides with focus stop Minimum fine focusing adjustment at 0 002mm range...

Page 13: ...0 x 480 0 3MP MS Windows Mac OS X 10 6 10 8 Included A3513U 1280 x 1024 1 3MP MS Windows Mac OS 10 8 and up Linux 2 6 or above A3520U 1600 x 1200 2 0MP A3530U 2048 x 1536 3 0MP A3550U 2592 x 1944 5 0M...

Page 14: ...is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down specimen at the bottom Turn slide over so the cover glass faces up Diopter adjustment is not set pr...

Reviews: