IMAGE PROBLEMS

 

Problem

 

Cause

Solution

 

Image moves while 
focusing 

Specimen rises from stage 
surface 

Secure the specimen in the slide 
holder 

Revolving nosepiece is not in the 
click-stop position 

Revolve the nosepiece to the 
click-stop position 

Image tinged yellow 

Blue filter not used 

Use daylight blue filter 

Lamp intensity is too low 

Adjust the light intensity by rotating 
the intensity control dial 

Image is too bright 

Lamp intensity is too high 

Adjust the light intensity by rotating 
the intensity control dial 

Insufficient 
brightness 

Lamp intensity is too low 

Adjust the light intensity by rotating 
the intensity control dial 

Aperture diaphragm closed too far  Open to the proper setting 

Condenser position too low 

Position the condenser upward 

 

 

MECHANICAL PROBLEMS

 

Problem

 

Cause

Solution

 

Image will not focus 
with high power 
objectives 

Slide upside down   

Turn the slide over so the cover 
glass faces up 

Cover glass is too thick 

Use a 0.17mm cover glass 

High power objective 
contacts slide when 
changed from low 
power objective 

Slide upside down 

Turn the slide over so the cover 
glass faces up   

  Cover glass is too thick   

Use a 0.17mm cover glass 

Diopter adjustment is not set 
properly 

Readjust the diopter settings 

Slippage of focus 
when using the 
coarse focusing 
knob Fine focus is 
ineffective 

Tension adjustment is set too low 

Increase the tension on the focusing 
knobs 

Tension adjustment is set too high

Loosen the tension on the focusing 
knobs 

 

Summary of Contents for M826

Page 1: ...User Manual Binocular Biological Microscope Model M829 MicroscopeNet com...

Page 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Installation 3 2 Operation 5 3 Specifications 7 4 Troubleshooting Guide 8...

Page 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...ently for dirt or oil clean if necessary Use the 10x objective to initially focus the image of your specimen When changing the objective magnification rotate the objective nosepiece until you hear a c...

Page 6: ...the thumb screw 1 7 Installing the mirror optional your model may not have one 1 Turn the power off and unplug the power cord 2 Screw off the light cover on the base 3 Thread the black disc onto the...

Page 7: ...condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives Note The centering of the c...

Page 8: ...sharp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 2 8 Adjusting tension knob The tightness of t...

Page 9: ...e layer mechanical stage area 140mmx140mm movement range 75mmx50mm Illumination Halogen lamp 6V 20W Eyepieces Designation Magnification Field of View Focal Length Wide Field 10X 18mm 24 94mm Plan Fiel...

Page 10: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Page 11: ...sity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...

Reviews: