OPTICAL PROBLEMS 

Problem

 

Cause

 

Solution

 

Darkness at the 
periphery or 
uneven brightness 
in the field of view   

Revolving nosepiece not in click 
stop position   

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path   

The bulb of light source is not at 
the center   

Adjust the position of the bulb if 
possible  

Dirt or dust on the 
view  

Dirt or dust on the eyepiece, 
condenser, objective, collector 
lens or specimen   

Clean the lens with a camera 
cleaning kit   

 

No slide cover attached to the 
slide  

Attach a 0.17mm slide cover   

 

Slide cover is too thick or thin   

Use a slide cover of the appropriate 
thickness (0.17mm)   

 

Slide may be upside down   

Turn slide over so the cover-glass 
faces up   

Poor image quality  Immersion oil is on a dry 

objective (especially the 40x)   

Check the objectives, clean if 
necessary  

No immersion oil used with 100x 
objective  

Use immersion oil   

Air bubbles in immersion oil   

Remove bubbles   

Condenser aperture is closed or 
open too much   

Open or close properly   

Condenser is positioned too low  Position the condenser upward   

 

 

ELECTRICAL PROBLEMS 

Problem

 

Cause

 

Solution

 

Lamp does not 
light when 
switched on   

No electrical power   

Check power cord connection   

Lamp bulb burnt out   

Replace bulb   

Fuse blown out   

Replace fuse   

 

Summary of Contents for M825

Page 1: ...User Manual Binocular Compound Microscope Model M825 MicroscopeNet com...

Page 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 5 4 Specifications 7 5 Troubleshooting Guide 8...

Page 3: ...ure or humidity and dusty environments Ensure that the microscope is located on a smooth level and firm surface 4 If any specimen solutions or other liquids splash onto the stage objective or any othe...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...ece Tube 3 Diopter Ring 4 Viewing Head 5 Slide Holder 6 Stage 7 Condenser 8 Light Collector 9 Set Screw 10 Nosepiece 11 Objective 12 Stand 13 Coarse Focus Knob 14 Fine Focus Knob 15 Translation Knobs...

Page 6: ...2 3 Installation of the objectives 1 Adjust the coarse focus knob until the mechanical stage is at its lowest position 2 Install the 4x objective into the nosepiece Then in a clock wise direction rot...

Page 7: ...tive turn the 100X objective in position and rotate the focus stop lever clock wise all the way to the end so that the 100X objective will not contact the specimen while the stage is adjusted to its h...

Page 8: ...m to 70 80 of the N A value of the respective objective will provide an image of acceptable quality If you want to observe the image of the iris diaphragm remove one eyepiece and look through the tube...

Page 9: ...ves Achromatic 4X 10X 40X spring 100X spring oil Condenser Abbe NA 1 25 w iris aperture diaphragm and filter holder Rack and pinion adjustment Focus Mechanism Coaxial coarse and fine focusing knobs on...

Page 10: ...d to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over so the cover glass faces up P...

Page 11: ...tensity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward IMAGE PROBLEMS Problem Cause Solution Image will not focus wi...

Reviews: