- 6 -

3

 

Operation 

3.1 Adjusting illumination 

1)  Connect the power cord and turn on the power switch (24).   

2)  Rotate the variable intensity dial (23) to increase or decrease the brightness. 

3.2 Placing specimen 

1)  Place the slide on the mechanical stage (8).   

2)  Use the slide holder (7) to gently secure the slide. Turn the X and Y adjustment knobs 

(22) to position the specimen for viewing. 

Caution:

 Be sure not to allow an objective to touch a specimen slide when changing 

objectives. 

3.3 Focusing 

1)  With the 10x objective in position, raise the mechanical stage using the coarse focus 

knob until the specimen is close to the objective 

2)  Turn the coarse focus knob until the specimen is in focus.   

3)  Then use the fine focus knob to obtain a sharp image. You may now switch to 

another magnification objective. 

Tips: 

To prevent your specimen slide from making contact with an objective, raise the 

stage to its highest position without contacting the 100x objective, then tighten the upper 
limit mechanical stage lever.

 

3.4 Condenser focusing 

Turn the condenser focus knob (10) to raise or lower the condenser. The condenser 
is raised when using high magnification objectives and lowered when using low 
magnification objectives. 

Note

:  

 

The centering of the condenser and the light axis of the objective are factory adjusted. 
Do not attempt re-adjustment. 

 

The highest position of the condenser has been factory adjusted. Do not attempt 
re-adjustment.

 

3.5  Adjusting iris diaphragm 

Slide the Iris diaphragm lever left or right to adjust the 
aperture size. 

Note: 

The iris diaphragm is designed to adjust the aperture 

size, not to adjust brightness. 

 

3.6  Adjusting interpupillary distance 

While observing with both eyes, hold the left and right eye 
tubes, slide in and out the tubes. The interpupillary distance is correct when the left 
and right fields of view converge completely into one image. 

3.7  Adjusting eyepiece diopter 

1)  Using the 10x objective and your right eye only, observe your specimen through the 

Iris Diaphragm Lever

Summary of Contents for M8244

Page 1: ...User Manual Dual Head Binocular Compound Microscope Model M8244 MicroscopeNet com...

Page 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 8 5 Troubleshooting Guide 9...

Page 3: ...ortant the lamp lamp housing and adjacent parts will become very hot Do not touch these parts until they have completely cooled Never attempt to handle a hot halogen bulb 6 All electrical connectors p...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...8 Mechanical Stage 9 Condenser 10 Condenser Focus Knob 11 Light Collector 12 Base 13 Viewing Head V 14 Thumb Screw V 15 Thumb Screw H 16 Viewing Head Block 17 Thumb Screw 18 Body 19 Tension Collar 20...

Page 6: ...eeding higher magnification objective Note When changing the objective magnification rotate the objective nosepiece until you hear a click sound This ensures the objective is centered in the optical l...

Page 7: ...pieces off and be certain that the head is securely locked by the thumb screw 2 7 Installing the mirror optional your model may not have one 1 Unplug the power cord 2 Screw off the light collector Len...

Page 8: ...highest position without contacting the 100x objective then tighten the upper limit mechanical stage lever 3 4 Condenser focusing Turn the condenser focus knob 10 to raise or lower the condenser The...

Page 9: ...harp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 3 8 Adjusting tension knob The tightness of th...

Page 10: ...nt Focus Mechanism Coaxial coarse and fine focusing knobs on both sides w limit lock Minimum fine focusing adjustment at 0 002mm range 28 mm Stage Large double layer mechanical stage Dimension 5 1 2 x...

Page 11: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Page 12: ...nsity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward Mechanical Problems Problem Cause Solution Image will not focus...

Reviews: