Omax M8244 User Manual Download Page 10

 

 

www.microscopenet.com

 

 

- 8 -

4

 

Specifications 

General

 

Model M8244 

Total Magnification 

40X, 64X, 100X, 160X, 400X, 640X, 1000X, 1600X 

Viewing Head 

Dual viewing head, binocular, inclined 45º, swiveling 360º 
Interpupillary distance 55-75mm 
Adjustable diopter on both eyepiece tubes   

Eyepieces 

2 pair of wide field eyepieces WF10X/18 

Objective Tube Length 

160mm 

Nosepiece Revolving 

quadruple 

Objectives 

Achromatic 4X, 10X, 40X (spring), 100X (spring, oil) 

Condenser 

Abbe, NA=1.25, w/ iris diaphragm and filter holder 
Rack and pinion adjustment 

Focus Mechanism 

Coaxial coarse and fine focusing knobs on both sides w/ limit lock 
Minimum fine focusing adjustment at 0.002mm, range 28 mm 

Stage 

Large double layer mechanical stage 
Dimension: 5-1/2” x 5-1/2” (140mmx140mm) 
Movement range: 3” x 2” (75mm X 50mm) 

Illumination Transmitted: 

6V/20W, Halogen, Variable intensity 

Power Supply 

AC 110V,

 

60HZ 

Net Weight 

12.6 lb (5.7 kg) 

 

Eyepieces 

Designation 

Magnification 

Field of View 

Focal Length 

Wide Field 

10X 

18mm 

24.94mm 

 

Objectives

 

   

Magnification 

Numerical 

Aperture 

Thickness of 

slide cover 

Focus 

Working 

Distance 

Working 

Mode 

4X 0.10 

0.17mm 

31.04mm 

37.5mm 

Dry 

10X 0.25 

0.17mm 

17.13mm 

7.316mm 

Dry 

40X(S) 0.65 

0.17mm 

4.65mm 

0.632mm 

Dry 

100X(S, Oil) 

1.25 

0.17mm 

2.906mm 

0.198mm 

Oil 

 

 

 

 

Summary of Contents for M8244

Page 1: ...User Manual Dual Head Binocular Compound Microscope Model M8244 MicroscopeNet com...

Page 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Component Illustration 3 2 Installation 4 3 Operation 6 4 Specifications 8 5 Troubleshooting Guide 9...

Page 3: ...ortant the lamp lamp housing and adjacent parts will become very hot Do not touch these parts until they have completely cooled Never attempt to handle a hot halogen bulb 6 All electrical connectors p...

Page 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Page 5: ...8 Mechanical Stage 9 Condenser 10 Condenser Focus Knob 11 Light Collector 12 Base 13 Viewing Head V 14 Thumb Screw V 15 Thumb Screw H 16 Viewing Head Block 17 Thumb Screw 18 Body 19 Tension Collar 20...

Page 6: ...eeding higher magnification objective Note When changing the objective magnification rotate the objective nosepiece until you hear a click sound This ensures the objective is centered in the optical l...

Page 7: ...pieces off and be certain that the head is securely locked by the thumb screw 2 7 Installing the mirror optional your model may not have one 1 Unplug the power cord 2 Screw off the light collector Len...

Page 8: ...highest position without contacting the 100x objective then tighten the upper limit mechanical stage lever 3 4 Condenser focusing Turn the condenser focus knob 10 to raise or lower the condenser The...

Page 9: ...harp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 3 8 Adjusting tension knob The tightness of th...

Page 10: ...nt Focus Mechanism Coaxial coarse and fine focusing knobs on both sides w limit lock Minimum fine focusing adjustment at 0 002mm range 28 mm Stage Large double layer mechanical stage Dimension 5 1 2 x...

Page 11: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Page 12: ...nsity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward Mechanical Problems Problem Cause Solution Image will not focus...

Reviews: