Quick Reference Guide to Near Infra-red Imaging with VISERA ELITE II

Disclaimer:

 Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each 

item of equipment which should be thoroughly read and understood prior to handling, use and storage of the equipment. 

Troubleshooting

Camera Head

Controlling NIR Imaging

Through the NIR light source.

Or select from the camera head programmable buttons.

Press the Optical Digital mode on and toggle the mode 
select until IR is selected.

No power

• 

Ensure everything is switched on.

• 

Check the isolation switch on the transformer is 
turned on.

No image

• 

Ensure everything is switched on including the 
monitor which can be switched off by accident.

• 

Unplug camera head, colour bars should appear. If no 
colours appear, contact Olympus Service Helpdesk.

IR Observation is disabled (not illuminated)

• 

Check the telescope is an Olympus IR compatible  
one with an IR marking on the barrel.

• 

Check the light guide cable is connected to  
the IR light source and remove any cable from  
the processor.

The endoscopic image is completely green

• 

Check an IR compatible telescope is being used.

• 

Turn down the theatre lights as far as possible.

The three programmable buttons are assigned as follows:

1.

 IR – on and off

2.

 Colour mode (overlay (mode 1) or pure IR (mode 2)

3. 

Record/Pause

Control panel

Observation 

mode select 

button

Observation 

mode selection 

indicators

Observation 

mode 

indicator

Observation 

mode button

 

Issue 1 

05/02/2021

 

KMF-08928 

NCDR 02866