Chapter 3 Preparation and Inspection
OSF TYPE V60 OPERATION MANUAL
• Only Air/Water Valve (MB-196) and Suction Valve (MB-197)
should be used with this endoscope. These valves are not
designed for use with models other than this scope and
10/20/30/100/130/200/230 Series endoscopes.
Inspection of the Biopsy Valve
The Biopsy Valve is a consumable item that should be
inspected before each use. Replace it with a new one if
irregularities are observed by following inspection. An
irregular, abnormal or damaged valve can reduce the efficacy
of the endoscope’s suction system, and may leak or spray
patient debris or fluids, posing an infection-control risk.
1.
Confirm that the slit and hole on the Biopsy Valves have no splits, cracks,
deformation, discoloration or other damage (see Figure 3.6).
Figure 3.6
2.
Attach the cap to the main body as shown in Figure 3.7.
Figure 3.7
Slit
Cap
Main body
Hole
Splits, Cracks
Discoloration
Normal
Abnormal
Slit
Main body
Cap