background image

 

PV9353F4E007 15 

(14) Check 

the 

result. 

-> See 

4.15 Checking the results

(15)  Shut down the software, and then shut down the controller supply unit and computer. 

-> See 

4.16 Quit

 

4.2. Start-up 

 
 
 
 
 
 
 
 
(1)  Turn on the power switch of the controller supply unit. 
 
(2)  Turn on the "LAMP2" switch of the controller supply unit. 
 
(3)  Start the computer, and then double-click the shortcut icon       on the desktop to 

start the program. 

 
 

4.3. Revolving nosepiece settings 

(1)  Configure the revolving nosepiece settings. 

When executing a 45-degree reflection measurement using the Reflectivity 
Measurement Unit for 45-Degree, adjust the revolving nosepiece position to the 
through position (the position where the objective lens is not set). 
 
 
 

 

 

 

 
 
 
 

 
 
 

For computer operation, refer to the instruction manual of the 
computer and use it in a proper manner. 

It takes approximately 15 minutes until the halogen lamp steadily 
operates after the lamp switch is turned on. Execute warm-up as 
necessary. 

Revolving nosepiece   

Summary of Contents for USPM-45EXP

Page 1: ...sure to read this instruction manual before use Since the description of this document is limited to contents related to the 45 degree expansion unit please also read the separate NIR Micro Spectropho...

Page 2: ...of this Product 6 1 Checking Delivered Items 7 2 Part Names and Functions Specifications 9 3 Installation and Connection 12 4 Operations 14 5 Notes on Storing 24 6 If Abnormality Occurs 24 7 Conformi...

Page 3: ...aler or our sales department for replacement or other inquiries Marks Meanings Do not touch because the surface is hot Otherwise you can get burned This indicates a warning or caution covering a wide...

Page 4: ...this product are as follows Indicates the proper upper direction of the packaged cargo Indicates that the packaged cargo must not be exposed to rain Indicates that packaged cargo must be carefully han...

Page 5: ...question regarding the content of this operation manual please contact your dealer or our sales department Prohibition of repairs or modifications to the product Do not disassemble or modify this pro...

Page 6: ...g Strictly observe the following when handling this product In addition observe the warnings or cautions described in each chapter Gently operate each adjustment knob Gently turn the adjustment knobs...

Page 7: ...e product free of charge 2 This guarantee is not applied to any failure that has been caused by another reason for example a failure in a peripheral device or a problem in the installation location en...

Page 8: ...you should notice any mistake or omission regarding a description please inquire to the contact point listed in the attached sheet Notwithstanding the foregoing we cannot be held responsible for any...

Page 9: ...r dealer or a branch of our company It is a reference sample of 45 degrees for the reflection measurement Reference data CD is attached Because the mirror court is given I will recommend the update of...

Page 10: ...included 1 45 degree reflection measurement reference sample collimator unit for transmittance tilt stage package Name Quantity Remarks 5 collimator unit for transmittance See Note 1 1 6 Tilt stage S...

Page 11: ...ce Light source unit for optional transmission measurement or 45 degree measurement to be used by fixing the lamp house using the fixing knobs two locations For a detailed explanation see 3 Installati...

Page 12: ...larization unit Adjust the polarization condition between the P polarization S polarization and random polarization by using the polarization switching knob 5 Tilt stage Adjust the two axis tilt again...

Page 13: ...ree Model name USPM 45EXP Wavelength range 380 1 050 nm Wavelength resolution Approximately 5 nm wavelength indication resolution 1 nm Repeatability 3 0 3 or less 430 1 010 nm 1 0 or less in other ran...

Page 14: ...se into the original position The lamp connector must be positioned to the right of the device Check the portion surrounded by a circle below and check the screw hole position and the lamp house fixin...

Page 15: ...not connected the intended safety performance cannot be ensured 5 Protection key Any USB port on the computer For cable setting keep cables away from something sharp Be careful not to apply excessive...

Page 16: ...onfigure the work setting file See 4 6 Setting up the work configuration file For the reference settings select FILE and select the file in the attached 45 degree reference sample data CD 6 Set the su...

Page 17: ...tart the program 4 3 Revolving nosepiece settings 1 Configure the revolving nosepiece settings When executing a 45 degree reflection measurement using the Reflectivity Measurement Unit for 45 Degree a...

Page 18: ...direction is the direction against the stage surface Refer to the figure below Far position Random polarization Intermediate position P polarization Near position S polarization Polarization switching...

Page 19: ...tion 1 Choose Setup Work on the File menu The work setup window will open 2 Configure the reference settings 1 Click and select FILE 2 Click the portion surrounded by a circle and open the rf file rf...

Page 20: ...e FOCUS window and the CT window by double clicking the black portion of the LIVE window The FOCUS window or the CT window can be selected from the file menu Live The FOCUS window is displayed when th...

Page 21: ...ue of X Y shown at the lower left in the FOCUS window becomes the smallest For a more precise measurement it is recommended that you adjust the value of X Y to X 0 1 Y 0 1 Tilt not in the correct posi...

Page 22: ...ht amount of the reference light flux reference does not exceed the maximum received light amount of the light receiving element Follow the procedures below to adjust the sampling interval 1 Click SMP...

Page 23: ...ion panel 4 12 Setting a sample to be measured 1 Set the sample on the top surface of the tilt stage Locate the sample in the hole at the center of the stage 4 13 Adjusting the position of a sample to...

Page 24: ...urement result 3 DIAGRAM window The measurement result of the chromaticity diagram is displayed The chromaticity diagram can show the XY chromaticity diagram and Lab chromaticity diagram By double cli...

Page 25: ...the lower limit of the spectroscopic quantity defined in the system configuration window 6 LUMINANCE TEXT window The spectroscopic light amounts of the background reference or measurement are displaye...

Page 26: ...n Manual 5 3 Wavelength offset 6 If Abnormality Occurs Also check NIR Micro Spectrophotometer USPM W B Operation Manual 6 Troubleshooting 7 Conforming to Laws and Regulations Refer to NIR Micro Spectr...

Page 27: ...Adjust the revolving nosepiece position to the through position the position where the objective lens is not set 4 Set the BK7 reference sample on the tilt stage 5 Adjust the tilt of the two axes to...

Page 28: ...s reserved The contents of this manual shall not be reproduced or distributed in part or in whole without Olympus s consent The company names and product names referred to in this document may be the...

Page 29: ...us Tower 446 Bongeunsa ro Gangnam gu Seoul Korea 135 509 Tel 82 2 6255 3458 OLYMPUS THAILAND CO LTD 159 Serm mit Tower 9th floor unit 906 910 Sukhumvit21 road North Klongtoey Wattana Bankok 10110 Thai...

Reviews: