background image

This sheet is for quick reference only. Please refer to the relevant User’s Manual(s) for 
instructions, warnings and cautions. Nothing in this presentation is meant to supersede  
or replace the instructions for use applicable to each specific device, or the processes  
and procedures in place at your facility.

USG-410 – Quick Reference Guide

Troubleshooting

Error

Possible causes

Solution

U504/U601  
Probe damage

Instrument has not been replaced.

Replace the instrument.

Plug is not connected properly to the socket.

Connect plug and socket firmly.

Transducer temperature is too high.

Stop output until the temperature cools down.

Transducer malfunction or probe damage/malfunction.

Replace the transducer/instrument.

U508/U511  
Short circuit 
(THUNDERBEAT only)

Output is generated while a metallic object is grasped.

Confirm that no metal object is grasped between jaw and probe tip.

Output in body fluids.

Remove excess fluids.

Grasping section wear out.

Replace instrument.

U509  
Ultrasonic level error

Excessive load is applied to the distal end. 

Reduce the force applied to the distal end or to the probe tip.  
Withdraw body fluid and tissue on the grasping section, probe tip, and shaft surface.

Transducer plug is not connected properly to the THUNDERBEAT / SONICBEAT 
socket at USG-410. 

Confirm that the transducer plug is connected to the THUNDERBEAT / SONICBEAT socket at USG-410.  
If not, connect it properly. 

Transducer malfunction or probe damage/malfunction.

Replace the transducer/instrument.

U510  
Ultrasonic frequency  
error

Connection of transducer and instrument is loose.

Connect transducer firmly with instrument.

Excessive load is applied to the distal end.

Reduce the force applied. Remove body fluid and tissue on the grasping section, probe tip, and shaft surface.

Transducer socket and plug are not connected properly.

Check if plug and socket are connected properly.

Transducer temperature is too high.

Stop output until the temperature cools down.

Transducer malfunction/probe damage or malfunction.

Replace transducer/instrument.

U512  
Open circuit  
(THUNDERBEAT only)

Instrument is activated without holding tissue.

Re-grasp tissue and confirm activation.

Transducer and instrument or transducer socket and plug are not connected properly. Check if transducer and instrument or plug and socket are connected properly.

Transducer/instrument malfunction.

Replace transducer/instrument.

U514  
Seal incomplete  
(THUNDERBEAT only)

The specified output time period has expired (maximum output time: 15 s).

Withdraw tissue or fluids from jaw and probe tip. Re-grasp tissue and confirm activation. 

W7083361_00 · 12/22 · GMA

Reviews: