USG-410
Get started
1. Connect THUNDERBEAT or SONICBEAT via a transducer to USG-410.
- Align the symbols on plug and socket and fully insert the transducer
plug in the socket of USG-410.
- The USG-410 recognizes the connected instrument and displays the
corresponding set screen (see next page).
2. Verify the ultrasonic output.
3. Verify the RF bipolar output (THUNDERBEAT only).
Power switch
Transducer socket
Touch screen
Pre-Procedure Device Checks
The activation test for ultrasonic output needs to be done every time
a THUNDERBEAT or SONICBEAT instrument is connected.
Activation test for ultrasonic output
1. Connect THUNDERBEAT or SONICBEAT and tap the “Test” button.
Caution: Do not apply the instrument on the patient during activation test.
2. Activate instrument with the jaws opened following the instructions on the screen.
3. When the activation test result is displayed, release the hand switch and confi rm
with OK.
4. After a successful test, the instrument can be used safely.
Verifi cation of the RF bipolar output (THUNDERBEAT only)
1. Fill a sterile container with saline.
2. Open the THUNDERBEAT instrument jaws and submerge only the distal half
of both jaws in saline.
3. Press and hold the SEAL button (blue).
4. Confi rm the following during the SEAL
mode output:
- The output screen is displayed.
- The output tone sounds.
5. After the confi rmation, release the SEAL
button.
6. Wipe any remaining saline off the jaws
with a sterile, lint-free cloth.