Chapter 6 Care, Storage and Disposal
57
HIGH FLOW INSUFFLATION UNIT UHI-3
Steam sterilize (autoclave) within the parameters given in Table 6.1 below. When
steam sterilizing, follow the hospital’s protocol and the sterilization equipment
manufacturer’s instructions while adhering to the limits given here.
6.4
Cleaning of the UHI-3 main body, foot switch,
cylinder hose and medical gas pipeline adaptor
• Set the power switch to OFF and disconnect the power cord
before cleaning.
• Alcohol is not a sterilant or high-level disinfectant.
• The equipment is non-immersible. Do not submerge or allow
fluid to enter them
• Do not autoclave or gas sterilize the equipment. Doing so will
cause equipment damage.
• Never expose electrical contact points (system connector,
foot switch connector, and AC inlet) to fluids, as this will
impede contact.
1.
Should the equipment become soiled with blood or other potentially
infectious materials, first wipe off all gross debris and then decontaminate
the equipment using a surface disinfectant. Always follow the
manufacturer’s recommendations for surface disinfection. Make sure that
the equipment is completely dry before use.
2.
The equipment should be routinely cleaned. To remove dust, dirt and other
non-patient debris, wipe using a soft, lint-free cloth moistened with 70%
ethyl or isopropyl alcohol.
Sterilizer
Temperature
Exposure time
Prevacuum
132 – 134
°
C (270 – 274
°
F)
5 minutes
Table 6.1