DMTA041-01EN [U8778479], Rev. C, May 2014
Maintenance and Troubleshooting
37
4.3.1
Cleaning the Fan Filters
To clean the fan filters
1.
Unscrew and remove the four knurled screws that secure each fan filter (see
Figure 4-1 on page 37).
2.
Remove the fan filter.
3.
Using a compressed-air jet, clean the filter on both sides.
4.
Place the filter back in the appropriate position.
5.
Screw the four knurled screws back into place.
Figure 4-1 Fan filter removed from the TomoScan FOCUS LT
4.3.2
Changing the Fan Filters
To change the fan filters
1.
Unscrew and remove the four knurled screws that secure each fan filter (see
Figure 4-1 on page 37).
2.
Remove the fan filters.
3.
Change the fan filters (Olympus part no. 26AB0008 [U8780134]).
4.
Place the filters back in the appropriate positions.
Summary of Contents for TomoScan FOCUS LT
Page 6: ...DMTA041 01EN U8778479 Rev C May 2014 Table of Contents vi...
Page 10: ...DMTA041 01EN U8778479 Rev C May 2014 Labels and Symbols 4...
Page 18: ...DMTA041 01EN U8778479 Rev C May 2014 Important Information Please Read Before Use 12...
Page 30: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 1 24...
Page 40: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 3 34...
Page 64: ...DMTA041 01EN U8778479 Rev C May 2014 Chapter 5 58...
Page 74: ...DMTA041 01EN U8778479 Rev C May 2014 List of Figures 68...
Page 76: ...DMTA041 01EN U8778479 Rev C May 2014 List of Tables 70...
Page 82: ...DMTA041 01EN U8778479 Rev C May 2014 Index 76...