![Olympus OmniScan SX User Manual Download Page 71](http://html1.mh-extra.com/html/olympus/omniscan-sx/omniscan-sx_user-manual_4296739071.webp)
DMTA-20043-01EN [U8778624], Rev. B, May 2013
Connector References 63
3
In
DIN1/
Preset1
Digital input 1/preset
axis 1.
Programmable input.
Can be configured as
generic input 1, or as
the preset of encoder 1.
Refer to the
OmniScan MXU
Software User’s Manual
(“Configuring the
Digital Input” section)
for information on
programming this
input.
To preset, you must use
a high-level signal with
a minimum signal
length of 50 ms.
N/A
TTL
4
In
DIN2/
Preset2
Digital input 2/preset
axis 2.
Programmable input.
Can be configured as
generic input 2, or as
the preset of encoder 2.
Refer to the
OmniScan MXU
Software User’s Manual
(“Configuring the
Digital Input” section)
for information on
programming this
input.
To preset, you must use
a high-level signal with
a minimum signal
length of 50 ms.
N/A
TTL
Table 12 Pinout for the scanner interface LEMO connector
(continued)
Pin
I/O
Signal
Description
Current
Level
Summary of Contents for OmniScan SX
Page 6: ...DMTA 20043 01EN U8778624 Rev B May 2013 vi Table of Contents ...
Page 8: ...DMTA 20043 01EN U8778624 Rev B May 2013 viii List of Abbreviations ...
Page 24: ...DMTA 20043 01EN U8778624 Rev B May 2013 16 Important Information Please Read Before Use ...
Page 26: ...DMTA 20043 01EN U8778624 Rev B May 2013 18 Introduction ...
Page 58: ...DMTA 20043 01EN U8778624 Rev B May 2013 50 Chapter 4 ...
Page 68: ...DMTA 20043 01EN U8778624 Rev B May 2013 60 Chapter 6 ...
Page 80: ...DMTA 20043 01EN U8778624 Rev B May 2013 72 List of Figures ...
Page 82: ...DMTA 20043 01EN U8778624 Rev B May 2013 74 List of Tables ...