Troubleshooting Guide
65
If this instrument is visibly damaged, does not function
as expected, or is found to have irregularities during the
procedures described in “Chapter 2 Installation and
Connection” and “Chapter 3 Inspection Before Use” in
“INSTRUCTIONS (BASIC)” and “INSTRUCTIONS
(ADVANCED),” do not use this instrument and contact
Olympus for repair according to “7.2 Sending this
Instrument for Repair” on page 73.
Some problems that appear to be malfunctions may be
correctable by referring to “7.1 Troubleshooting Guide”
below. If the problem cannot be resolved by the
described remedial action, stop using this instrument
and send it to Olympus for repair.
Olympus does not repair accessory parts. If an accessory
part becomes damaged, contact Olympus to purchase a
replacement.
WARNING
Never use this instrument if an abnormality is suspected.
Damage or irregularity in the instrument may
compromise patient or operator safety and may result in
more severe equipment damage.
7.1
Troubleshooting Guide
The following table shows the possible causes of and
countermeasures against problems that may occur due to
equipment setting errors or deterioration of consumable.
Problems or failures due to other causes than those listed
below should be serviced. As repair performed by
persons who are not qualified by Olympus could cause
patient or operator injury and/or equipment damage, be
sure to contact Olympus for repair as described in “7.2
Sending this Instrument for Repair”.
WARNING
If any irregularity is suspected with this instrument, first
turn it OFF and ON again.
Chapter 7 Troubleshooting
Irregularity
Possible cause
Solution
The power fails to come on. The power cord is not connected.
Connect to a wall mains socket outlet as described in “2.4
Connection to the AC Mains Power Supply” in
“INSTRUCTIONS (BASIC).”
The power cord is connected improperly.
The power cord is broken.
Replace the new power cord.
The mobile workstation is turned OFF.
Turn the mobile workstation ON.
This instrument is turned OFF.
Turn this instrument ON.
The PRINT lamp is kept
lit without going off
after setting the power
switch of this
instrument to ON.
A trouble occurred inside this instrument.
Turn this instrument OFF, then ON again after a while.
The PRINT lamp is lit
even when the
instrument is not
printing.
A trouble occurred inside this instrument.
Turn this instrument OFF, then ON again after a while.
The buzzer keeps
sounding.
A trouble occurred inside this instrument.
Turn this instrument OFF, then ON again after a while.