DMTA-10040-01EN, Rev. E, February 2018
Chapter 5
192
FULL NEXT key (
) four times and set the
RDG2 TYP
to
OFF
. When done,
press the Return key (
).
The list of dual frequency parameters is shown in Figure 5-99 on page 192.
Figure 5
‑
99 The list of dual frequency parameters
To calibrate the signals
1.
Press the MAIN FILTER menu key (
) once, and then set the
FREQ
(A key)
value to 3.0 kHz using the knob.
2.
Press the MAIN FILTER menu key (
) again, and then set the
FREQ 2
(A key)
value to 1.0 kHz.
3.
Place the probe on position 1 of the standard (see Figure 5-100 on page 193), and
then, while firmly pressing the probe on the standard, press the A-LIFT NULL
key (
).