Figure 5
‑
100 Position 1 of the standard
4.
Lift the probe into the air, then begin to gently tap the probe on position 1 of the
standard, and, while tapping, press the ANGLE key (
), and then adjust the
ANGLE
(for frequency 1) until the lift-off signal is oriented towards the right side
of the screen (the green signal in Figure 5-101 on page 193).
If necessary, you can press the ERASE key (
) to clear screen, which may
make it easier to adjust the
ANGLE
value.
Figure 5
‑
101 Adjusting the angle of the frequency 1 lift
‑
off signal
5.
While continuing to gently tap the probe on position 1, press the ANGLE key
(
) twice to open the
ANGLE 2
parameter, and then adjust the angle of the
frequency 2 lift-off signal until it also is oriented towards the right side of the
screen, superposed on the frequency 1 signal (see Figure 5-102 on page 194).
Summary of Contents for nortec 600
Page 8: ...DMTA 10040 01EN Rev E February 2018 Table of Contents viii...
Page 16: ...DMTA 10040 01EN Rev E February 2018 Labels and Symbols 6...
Page 30: ...DMTA 10040 01EN Rev E February 2018 Introduction 20...
Page 58: ...DMTA 10040 01EN Rev E February 2018 Chapter 1 48...
Page 71: ...DMTA 10040 01EN Rev E February 2018 Software User Interface 61 Press the Return key to exit...
Page 72: ...DMTA 10040 01EN Rev E February 2018 Chapter 2 62...
Page 342: ...DMTA 10040 01EN Rev E February 2018 Chapter 7 332...
Page 356: ...DMTA 10040 01EN Rev E February 2018 Appendix B 346...
Page 366: ...DMTA 10040 01EN Rev E February 2018 List of Figures 356...
Page 368: ...DMTA 10040 01EN Rev E February 2018 List of Tables 358...