DMTA017-01EN, Rev. F, August 2016
Chapter 4
34
5.
Using a flathead screwdriver, remove the screw (at the back of the SLA) that
secures the yoke to the spring-loaded arm.
6.
Remove the yoke, probe holder, and probe assembly from the scanner.
7.
Using a hexagonal key, remove the screw (on the side of the yoke) that holds the
parts together (see Figure 4-5 on page 34).
Figure 4
‑
5 The screw that secures the yoke assembly
8.
Using a cross-head screwdriver, loosen the four screws (see Figure 4-6 on
page 34), and then remove the probe from the wedge.
Figure 4
‑
6 The probe holder
9.
Install the new probe.