G3 WORKSTATION
Begin treatment:
a
Press and release the BLUE foot pedal OR press the MODE button (located in
the center of the controller with a round arrow) ONCE (This will set the system to RF
SENSING OFF).
B
Press and release the BLUE foot pedal OR press the Mode button a second time.
(This will set the system to RF SENSING ON, Impedance values are recognized here).
C
Press and release the BLUE foot pedal OR press the Mode button a third time.
(This initiates the energy delivery sequence).
Pause: Press the BLUE foot pedal
•
Restart: Press BLUE foot pedal once more
•
reset and additional treatment:
a
After each completed lesion, press the BLACK mode foot pedal once to reset the
system to RF SENSING OFF.
B
Press and release the BLUE foot pedal will set the system to RF SENSING ON
(Impedance values are recognized).
C
Press the BLUE foot pedal once more to initiate energy delivery to the additional
site.
D
Audible tone is recognized every 50 Joules.
Settings using tCrF for turbinate:
Temperature: 85ºC
•
Joules: 350-500
•
Rapid lesion mode:
Temperature: Set to Rapid Lesion Mode
•
Joules: Single channel instruments: 300 peak
•
Impedance: 110 - 160 OHMS
•