Instructions Guide for G3 WorkStation 
Procedure Setup

G3 WorkStation

ENT

E0499049_Folder_Gyrus_G3_A5_2.indd   1

22.03.2010   12:59:18 Uhr

Summary of Contents for G3

Page 1: ...Instructions Guide for G3 WorkStation Procedure Setup G3 WorkStation ENT ...

Page 2: ...e Somnoplasty procedures As a result one technology platform can be used for a growing variety of surgical procedures providing enhanced value and versatility now and in the future G3 WorkStation 1 2 4 3 6 7 5 1 Return Electrode Connection 2 TCRF Handpiece Cable Connection 3 TCRF Set Target Temperature PlasmaCision PlasmaBlend Settings 4 Standby Power Button 5 Mode Menu 6 TCRF Set Target 7 PlasmaC...

Page 3: ...ate tonsil and dual channel tongue base and the gray cable for palate and single channel tongue base F Set the temperature using arrows on the left side of the controller G Set the Joules using arrows on the right side of the controller H See the charts below for temperature and Joules settings for each procedure Rapid Lesion Mode A Press the temperature arrow until 85ºC is reached B Release and p...

Page 4: ...energy delivery sequence Pause Press the BLUE foot pedal Restart Press BLUE foot pedal once more Reset and Additional treatment A After each completed lesion press the BLACK mode foot pedal once to reset the system to RF SENSING OFF B Press and release the BLUE foot pedal will set the system to RF SENSING ON Impedance values are recognized C Press the BLUE foot pedal once more to initiate energy d...

Page 5: ...perature 85ºC Joules Single channel instrument 600 750 per site Joules Dual channel instrument 1200 1500 per site Rapid lesion mode Temperature Set to Rapid Lesion Mode Joules Single channel instruments 300 peak Impedance 110 160 OHMS Settings using TCRF for base of tongue Temperature 85ºC Joules 800 1000 Rapid lesion mode Temperature Set to Rapid Lesion Mode Joules Dual channel instrument 600 pea...

Page 6: ...y the PlasmaKnife and set the appropriate default PlasmaBlend setting F To change the PlasmaBlend CUT COAG ratio press the BLACK mode foot pedal for 1 second until a beep is heard G Press the YELLOW pedal to increase the cut component in 5 increments or the BLUE pedal to increase the COAG component in 5 increments H Press the BLACK mode foot pedal to select the new PlasmaBlend ratio I Alternativel...

Page 7: ...ch for G3 WorkStation EG720100901 Powercord USA EG70353003 Tonsil PlasmaKnife 4 pack EG70353005 Dissector PlasmaKnife 4 pack EG70353006 jPlasmaKnife 5 pack EG1120411005 TCRF Turbinate Handpiece 5 pack EG1010211005 TCRF Palate Handpiece 5 pack EG2220221005 TCRF Dual Channel Tongue Base Handpiece 5 pack EG1200211005 TCRF Single Channel Tongue Base Handpiece 5 pack EG2420132805 TCRF Tonsil Handpiece ...

Page 8: ...ION Specifications design and accessories are subject to change without any notice or obligation on the part of the manufacturer Postbox 10 49 08 20034 Hamburg Germany www olympus europa com E0499049 3 000 03 10 e7 ...