Changing the halogen bulb

 

} The designated bulb model is the 12V100WHAL-L (PHILIPS 7724) or 

12V50WHAL-L (LIFE JC) halogen bulb.

Fully loosen the clamping screw 

 at the top of the halogen lamp 

housing using the Allen screwdriver provided with the system.

Lift up the halogen lamp housing 

2

 to remove it.

Tilt the bulb socket 

3

 by 90° in the direction of the arrow.

While pushing down the bulb clamping lever 

4

, hold the halogen 

bulb 

5

  with  gloves  or  a  piece  of  gauze,  insert  the  bulb  pins 

6

 

straight  and  fully  into  the  pin  position 

7

  on  the  lamp  socket.

Then return the lamp clamping lever gently back to the original position 
to clamp the bulb.

1

2

3

4

Do not touch the lamp directly.
If fingerprints, etc. are attached, wipe them off completely with 
the soft cloth, etc. to prevent the lamp from life reduction or 
burst.

CAUTION

Fit the halogen lamp housing from up and tighten the clamping screw 

1

 by applying downward pressure. 

5

1  Changing the lamp

•  The bulb, lamp housing and areas near these will be extremely hot during and right after use. 

Before replacing the bulb, exit the system and wait until the bulb and the lamp housing are fully 
cooled down. Then, replace to the appropriate bulb.

CAUTION

I  Changing the halogen bulb

2-2  Work related to the illumination

1

2

3

4

5

6

7

7

Summary of Contents for FLUOVIEW FVMPE-RS

Page 1: ...OVIEW FVMPE RS Multi Photon Laser Scanning Microscope Notice Thank you for your purchase of Olympus microscope at this time Retain this manual in an easily accessible place near a system for future re...

Page 2: ......

Page 3: ...service manual provided by Olympus The performance has influence on the feature of this system and there is a risk that unintended CLASS 4 laser light is emitted In order to maintain the full perform...

Page 4: ...trademarks of their respective owners Caution If the system is used in a manner not specified by this manual the safety of the user may be imperiled In addition the system may also be damaged Always...

Page 5: ...to the nosepiece 3 2 2 Work related to the illumination 7 1 Changing the halogen bulb 7 2 Changing the mercury lamp 9 2 3 Work related to optical unit 11 1 Fluorescence mirror unit 12 2 DM unit for th...

Page 6: ......

Page 7: ...piece caps and or nosepiece as this causes irradiation of laser beams that are parallel not diffused This system uses an IR pulse laser which has an element of danger to cause blindness if it enters t...

Page 8: ...erving record of it Other items necessary for preventing damages due to laser light Service work is work that can affect the performance of the device and such work is described in this service manual...

Page 9: ...g work CAUTION 2 1 Work related to the microscope When a nosepiece of a different type from the removed nosepiece is mounted it is necessary to change the nosepiece information settings for the touch...

Page 10: ...4 BX63LF FV30GF...

Page 11: ...d objective lens LUMPLFLN60XW LUMPLFLN40XW XLPLN25XWMP2 XLPLN25XSVMP2 XLSLPLN25XSVMP2 When using the objective lens marked with use the RMS adapter 4 provided with the piezo nosepiece adapter between...

Page 12: ...zo nosepiece 5 Connect the piezo nosepiece exclusive controller to FV30 ANALOG with the commercially available BNC cable Connect the control signal input connector CONTROL IN of the piezo nosepiece ex...

Page 13: ...the lamp socket Then return the lamp clamping lever gently back to the original position to clamp the bulb 1 2 3 4 Do not touch the lamp directly If fingerprints etc are attached wipe them off comple...

Page 14: ...he lamp housing c a b a b c d Remove the collector lens of the U LH100IR lamp housing a by loosening the three clamping screws b with an Allen wrench width across flats of 2 5 mm 1 While positioning t...

Page 15: ...g the lamp connector a Hold the black frame and raise the lamp slowly upward If the lamp is dropped the glass will break and you may be injured Pay careful attention when you replace the lamp Do not t...

Page 16: ...t method 1 30 Press the reset switch for two seconds or more 1 40 or later Press the reset switch and the lamp switch at the same time for five seconds or more 9 c c 9 9 Version 1 30 Version 1 40 or l...

Page 17: ...t U FDICTIR IX3 FDICT FV30 RFABXL DM unit for SIM scanner FV30 ADM800 FV30 ADMOPT FV30 HSUMP DM unit for External NDD Non Descanned Detector FV30 NDM690 FV30 NDM760 FV30 NDMVCOIR FV30 RFABXL Optical u...

Page 18: ...disabled and the optical performance of LSM could not be guaranteed Before attaching or detaching the fluorescence mirror unit be sure to engage the shutter in the light path Be sure to exit the FV30...

Page 19: ...o the space where the fluorescence mirror unit was removed until it touches the end 2 CAUTION 3 4 Hold the both right and left sides of the fluorescent mirror unit with your thumb and the index finger...

Page 20: ...HSUMP Attaching and removing the cover 1 Remove the cover of FV30 HSUMP by loosening the cover fixing screws four locations of FV30 HSUMP using the Allen screwdriver provided with the system Cover fix...

Page 21: ...mount dovetail of the DM into the mounting frame of FV30 HSUMP unit until it meets the end of the frame Tighten the screw securing the DM using the Allen screwdriver provided with the system 6 Loosen...

Page 22: ...RFABXL Attaching and removing the cover 1 Remove the cover of FV30 RFABXL by loosening the cover fixing screws four locations of FV30 RFABXL using the Allen screwdriver provided with the system Cover...

Page 23: ...the frame Tighten the screw securing the DM using the Allen screwdriver provided with the system DM for external NDD Screw securing the DM If the power of the system is turned ON check that the syste...

Page 24: ...he laser port FV30 LPADP When FV30 LPADP is placed to laser ports of FV30 HSUMP and FV30 SIMMP the secured cover can be removed with the screw M3 IR port FV30 SIMMP Vis port FV30 SIMMP IR port FV30 HS...

Page 25: ...less against the 8 hole of FV30 LPADP Tilt 90 0 6 against the surface to which FV30 LPADP is pushed The tolerable power density is the value when the incident position accuracy is satisfied If you ob...

Page 26: ......

Page 27: ......

Page 28: ...4 8F Olympus Tower 446 Bongeunsa ro Gangnam gu Seoul Korea 135 509 48 Woerd Avenue Waltham MA 02453 U S A Distributed by Shinjuku Monolith 2 3 1 Nishi Shinjuku Shinjuku ku Tokyo 163 0914 Japan Manufac...

Reviews: