Olympus FLUOVIEW FV300 Overview Download Page 6

5

X -t

SlantLine-t

FreeLine-t

Y -t

Y-Z-t

X-Y-t

ZoomIn-t

X-Z-t

X-Y-t

X-Y-Z-t

ZoomIn-Z-t

Patch clamp

BX61WI fixed stage upright mictranslation stage

Image acquisition

ROI designation

Intensity versus time
measurement

Calcium sparks in isolated cardiac myocyte

Dr. Sandor Gyorke
Texas Technical University

X -t

X -t

Time Course

Using different scanning modes to chart time-lapse changes efficiently.

High-speed (4 frames/sec) image
acquisition

For the high speed observation of the
sample, Fluoview is capable of scanning 4
frames per second in a fast scanning mode
at an image size of 512X512. By limiting the
image size, the frame rate will be even
faster. This scanning mode is suitable for
living cell observation.

Versatile line scanning modes have
many uses

The wide variety of the line scanning modes
(linear/slant/free-line) enables flexible
analysis of rapid time-lapse experiments.

Superior slice patching system

In combination with the unique fixed stage &
nosepiece focusing BX61WI microscope,
the FV300 provides a highly effective
system for slice patching. This unique set-
up has a small footprint for increased room
in a space-limited cage. The remote control
microscope options minimize the danger of
accidentally touching the delicate
experimental settings.Olympus also offers
ideal non-cover glass long working distance
water immersion objectives and an optional
XY translation stage that moves the entire
confocal microscope system while the
sample and other experimental hardware
remains in a fixed position.

Highly precise time-lapse analysis

Fluoview’s wide dynamic range of 12-bit or
4096 grey levels provides enough sensitivity
to detect even the slightest changes in
intensity. The user can designate multiple
regions of interest (ROI) by using drawing
tools. The fluorescence intensity or the ratio
can be analyzed with the intuitive GUI driven
program.

Calcium wave in Xenopus oocyte, Calcium Green staining, fluorescence
pseudo-colored fluorescence image after injection of inositol 3-trisphospate

Japan Science and Technology Corporation, Exploratory Research for Advanced Technology,
Mikoshiba cell control project, Prof. Aya Muto

Calcium wave in isolated cardiac myocyte

Dr. Sandor Gyorke
Texas Technical University

Long working distance, non-cover
glass water immersion objective

Immersion-type LUMPLFL objectives

The 40X water immersion objective in this series has a
3.3mm working distance and an extremely fine tip which
is suitable for micromanipulation using a fixed stage
upright microscope. It has a large N.A. (0.8) and is also
ideal for confocal observations. When using the BX61WI
fixed stage & nosepiece focusing upright microscope
with water immersion objectives, confocal imaging can
be used to monitor time-lapse fluorescence changes in
thick specimens such as brain slices.

Summary of Contents for FLUOVIEW FV300

Page 1: ...CONFOCAL LASER SCANNING BIOLOGICAL MICROSCOPES FV300 FLUOVIEW...

Page 2: ...stage upright microscope platforms the FV300 offers excellent versatility as well as top class laser scanning performance Highest image quality 12 bit 2048x2048 pixel resolution with economical cost E...

Page 3: ...r introduction w Beam collimator e Polarizer r Dichromatic mirror t Excitation dichromatic mirror y XY galvanometer mirror scanners u Pupil lens i Collector Lens o Pinhole turret 0 Emission beam split...

Page 4: ...ine scanning Intensity changes may be measured over a given period of time along the length of a freely drawn line such as the trace of an axon or along a cellular junction Clip scanning By cropping t...

Page 5: ...well Once optimized steps q r can be performed easily Scanning unit set up monitor display Human colon crypt Nuclei Blue TO PRO 3 Actin Green Alexa 488 APC gene product Red Alexa 568 Christine Anderso...

Page 6: ...er glass long working distance water immersion objectives and an optional XY translation stage that moves the entire confocal microscope system while the sample and other experimental hardware remains...

Page 7: ...low chart optional time course software TIEMPO Hardware and software support to optimize the environment for FRET FRET Input output of external trigger signal The optional time course software gives c...

Page 8: ...by describing the individual steps or phases within the experiment Users can specify detailed conditions and parameters for each step This function enables users to construct complex experimental prot...

Page 9: ...h point such as XYZ coordinates the Z axis acquisition range and the detector sensitivity 2 Up to 5 X 5 adjacent fields of view registered automatically Since adjacent fields of view are registered au...

Page 10: ...on driving mechanism that enables 0 01 m step control within the BX61 BX61WI and IX81 motorized microscopes high quality continuous cross sectional images can be acquired The 3D function also provides...

Page 11: ...hysioview software Axon PC side Working conditions Clampex Ver 8 0 or later Physiolink software FV300 side Colocalization image white Colocalization By using this function to analyze multi color speci...

Page 12: ...g pseudo colors Mituyo Kinoshita Pr Kentaro Arikawa Laboratory of Neuroethology Graduate School of Integrated Science Yokohama City University Purkinje cell in the rat cerebellum FITC vesicular GABA t...

Page 13: ...ish embryo Extended focus image of 10 mx28 slice Pr Yasuhiro Kamei Pr Shunsuke Yuba Institute for Molecular and Cellular Biology Osaka University GFP labeling of Drosophila adult brain with staining o...

Page 14: ...BA625IF BA660IF BA585 640 440 LD 405 LD 458 Multi Argon 488 Multi Argon 515 Multi Argon 543 Green HeNe 568 Krypton Argon 633 Red HeNe 488 Argon SDM515 SDM505 SDM570 RDM630 SDM630 SDM560 SDM600 ECFP E...

Page 15: ...lication Real time image calculation Kalman filtering peak integration software Hardware control Laser scanning unit microscope Each image display Single channel side by side merge cropping tiling ser...

Page 16: ...Multi Ar HeNe Red Green lasers Installation stand is not included in the unit Please consult your Olympus dealer for additional laser combinations External transmitted light detector system Fluorescen...