background image

7.

Confocal Aperture, 4.10

 – Controls the diameter of the pinhole.

8.

Kalman Averaging, 4.11 – 

Allows averaging signals from multiple scans to reduce noise.

9.

Information, 4.12 – 

Opens a window displaying the current pixel size.

Focus X2 and Focus X4 settings switch to 2 

µ

s dwell time. 

 These scan settings reduce resolution in 

Y axis by 2-fold and 4-fold, respectively. Use these settings with the AutoHV switch selected.

2.3  Settings for the fluorescence detectors.

Each fluorophore selected from the Dye List will activate one of the photomultiplier tube (PMT) 
channels and set its appropriate filters.  Three settings control the sensitivity and signal to noise ratio for 
each detector.  Refer to “Optimizing the Image” for details on these settings.

1.

HV, 5.1

 – The primary adjustment for the PMT.  HV controls the detector sensitivity and 

amplification.  

a. try to stay below 750, higher settings will introduce detector noise in the image
b. higher HV may used with averaged images to allow reduced laser intensity

2.

Gain, 5.2

 – This multiplies the analog electrical signal created by the PMT by a constant, before 

it is digitized.  While this can provide ‘brighter’ images, the noise and background intensities are 
increased as well.

a. leave set to “1” unless working with very low HV settings.

3.

Offset, 5.3

 – The Offset will raise or lower all pixel intensities to control the “Black Level” of 

the image.  It should be set so that the darkest values in the image are above zero.  

a. leave set to default values of approximately 5 to 7
b. set the display to gray scale, pixels with intensity = 0 will be pseudocolored blue
c. set the Offset so that no blue pixels ( intensity = 0) are present

2.4  Additional detector controls.

1.

Laser power level, 5.4

 - changing from this window may be more convenient than scrolling over 

to the laser control window.  

2.

Active detector check mark, 5.7

 – removing this check mark will turn off the channel and its 

associated laser line.

3.

Group number, 5.5

 - allows channels to be grouped for sequential channel acquisition. 

a. channels that do not share bleedthrough, e.g. Alexa488 and Cy5, they can be combined 

into the same group.  

b. channels in the same group are scanned simultaneously to reduce the amount of time 

required for sequential acquisition

Olympus Fluoview-1000 User’s Guide

V.M. Bloedel Hearing Research Center, Core for Communication Research

Center on Human Development and Disability, Digital Microscopy Center

May 11, 2011 

13

Summary of Contents for Fluoview-1000

Page 1: ...7 4 15 Line Settings Figure 7 5 15 Scan Controls 16 Scan Mode Figure 6 1 16 Dwell Time Figure 6 2 16 Scan Regions Figure 6 3 17 Image Aspect Ratio Figure 6 4 17 Box Size 6 5 17 Box Size effect on res...

Page 2: ...Macro 28 Image 5D plugin 28 File Import from the OIF Files directory 29 File Transfer 29 File transfer to Mac Pro Fluoview confocal file server 29 Saving files directly to Mac Pro Fluoview 29 USB and...

Page 3: ...el Size m versus Box Size and Magnification at Zoom 1 39 Microscope controls 41 Focus 41 Fine Coarse Focus 41 Lowering the Objectives 41 Manual shutter 41 Differential Interference Contrast DIC 41 DIC...

Page 4: ...eries Control Window 24 Figure 11 Z Series Control Window 24 Figure 12 DIC and Focus Controls Lower Half of the Microscope 40 Figure 13 Condenser and DIC Controls 42 Figure 14 Tilting the condenser 43...

Page 5: ...he previous user B Shut down 1 Transfer files from your account See File Transfer Section 2 Exit from Fluoview 3 Ignore the warning about light coming through the microscope 4 Turn off power strip B 5...

Page 6: ...is usually better than brightfield for low contrast samples such as monolayers or picking out structures in complex specimens 5 Epi fluorescence is usually faster than confocal to find the fluorescent...

Page 7: ...touch Fig 3 1 14 Select mirrors fluorescence filter cubes Microscope Controller Window Mirror click on desired mirror Fig 3 3 see Table 1 15 Fine focus coarse focus press the F C button below each fo...

Page 8: ...ansmitted light with the Image Acquisition Control Window Figure 4 1 10 Switch to fluorescent imaging Remove the DIC analyzer from the light path for optimal confocal images Figure 2 Enlarged view of...

Page 9: ...switch to the last filter used 3 Select fluorescence filter if desired a Microscope Controller Window Mirror click on DAPI FITC or TRITC blue green or red fluorescence Fig 3 3 Table 1 4 Adjust focus a...

Page 10: ...nel is displayed in a separate window 4 Set to gray scale control h 5 Set scanning to simultaneous Sequential checkbox Fig 4 9 is cleared 6 Initiate scanning Image Acquisition Control Window Focus X2...

Page 11: ...ctions b double click on any selected dye to remove it 3 Scroll through the list of dyes in the lower window 4 Select each desired fluorophore by double clicking on the name a if your particular dye d...

Page 12: ...ing the stage b may be used for recording rapid fluorescence changes such as ion measurements 2 Focus X4 4 5 Scans alternate odd lines replicating the scanned lines to generate an image of 4X reduced...

Page 13: ...ile this can provide brighter images the noise and background intensities are increased as well a leave set to 1 unless working with very low HV settings 3 Offset 5 3 The Offset will raise or lower al...

Page 14: ...a check mark Figure 7 1 Clicking in these boxes will toggle lines between active and blocked This control blanks the lines in the scanhead it does not turn the laser itself on or off 2 5 2 Line availa...

Page 15: ...he slider 2 5 5 Line Settings Figure 7 5 As with other controls the laser power may be adjusted by clicking anywhere in the control area scrolling the thumbwheel 1 increments or by clicking on the bla...

Page 16: ...ent moving over each point to deliver more photons exciting the fluorescent label to generate signal However a longer dwell time will increase photobleaching The time required to scan a field is deter...

Page 17: ...Normal rectangle defines the scanned region as the selected box size The Polygon options allows user defined regions within the box These may be rectangular circular or irregular polygons The laser be...

Page 18: ...dwell time may cause your image to become saturated 1 Begin scanning with Focus X2 or Focus X4 Focus mode and find your sample 2 Adjust the laser and detector settings 3 Click on Auto HV to set the re...

Page 19: ...plished by clicking on the Zoom Reset then changing the zoom to the new value 2 7 6 Scan Offset 8 6 Move the Scan Area within the Field of View by either clicking on the scan area and dragging it or b...

Page 20: ...ng Controls 9 1 Sequential Line vs Frame 9 3 Channels 9 3 Kalman Averaging 9 2 Groups 9 2A 9 3A Grouped Channels Olympus Fluoview 1000 User s Guide V M Bloedel Hearing Research Center Core for Communi...

Page 21: ...nels If a laser line also significantly excites other fluorophores their emissions may overlap sufficiently to cause bleedthrough As always control specimens are important Simultaneous is faster than...

Page 22: ...g 1 Live Window is set to gray scale control h in order to observe saturated pixels 2 Set Live Window for tile view each detector is viewed in its own pane 3 Click on Sequential 4 Select Frame 5 Click...

Page 23: ...collected with an increment of focus between each image plane This is an automated process where the user sets the upper and lower limits of the focus travel with an interval for incrementing the focu...

Page 24: ...End 10 5 Set Focus to 0 11 2 Optimal z Step 11 4 of Slices 11 3 Focus Increment 11 5 Clear Start End 11 1 Set Focus Limits 11 7 Step the Center of Z Series 11 6 Step Focus Full Step 1 2 Step Olympus F...

Page 25: ...append is controlled by the box next to the button The red number below the lens icon in this window displays the current axial position You can simplify keeping track of focus distance by clicking o...

Page 26: ...gions of interest and preserve the original 12 bit 4096 image intensity values Each format has its benefits and drawbacks 5 1 1 The Olympus OIB format This format writes all images and meta data to a...

Page 27: ...the scale bar show up in ImageJ or Photoshop However this means saving your file under yet another format Adding scale bars in ImageJ or Photoshop is simple and directions can be found on our website...

Page 28: ...Olympus datasets into ImageJ 5 4 1 Loci tools Plugin for ImageJ The LOCI plugin will import about 80 file formats including OIB and OIF You will be presented with choices for merging or splitting cha...

Page 29: ...iles shortcut on the confocal desktop 4 Navigate to your user account and the Files directory located in Drive D 5 Drag your confocal files from the User files directory to your directory in the Mac P...

Page 30: ...e list of favorites b type in your server address name or TCP IP address c Browse the network 5 5 6 Using FTP to Windows computers 1 FTP transfer to your server from Fluoview Mac Pro by the Fetch appl...

Page 31: ...aching 3 Channel Offset should be adjusted to remove blue pixels usually 5 to 7 4 HV and laser power levels should avoid saturation red mask 5 Establish imaging conditions using monochrome display rat...

Page 32: ...bodies subjected to immuno absorbtion are also useful 6 3 2 Positive controls Known positive controls are also important Positive controls labeled for each single label are crucial when imaging specim...

Page 33: ...he channel s in which the bleedthrough is observed High excitation intensity will cause the low probability tails of an emission spectrum to reach detectable levels in adjacent channels for most combi...

Page 34: ...s by the square root of the number photons from the mean of the population This variation represents 1 standard deviation on each side of the mean number of photons This translates into the reality th...

Page 35: ...nsities that can not be attributed to specific or desired label signal Background reduces contrast and dynamic range interferes with computerized feature recognition and image processing Sources of ba...

Page 36: ...r spot and in doing so changes the sampling rate for the image Zoom provides additional resolution without the need to increase magnification as with a widefield microscope and camera 6 10 1 Zoom redu...

Page 37: ...Optimal Sampling um Pixel Size Z Step 4X 1 504 54 258 0 654 23 6 10X 40 0 601 8 56 0 261 3 72 20X 75 0 319 2 12 0 139 0 921 40X 1 30 Oil 0 181 0 948 0 079 0 412 60X 1 35 Oil 100X 1 4 Oil 0 168 0 778...

Page 38: ...2 Box size This term refers to the number of sampling points in the scanned field A box size of 512x512 means that the field will be divided into 512 samples along each of 512 lines If the lines in t...

Page 39: ...0 132 2048x2048 0 621 0 310 0 155 0 103 4096x4096 0 310 0 155 0 103 0 051 The size of a pixel in a zoomed image is obtained by dividing the value from the above table by the zoom factor Shifting from...

Page 40: ...or for 2X oversampling Pixel size 2 wavelength NA Figure 12 DIC and Focus Controls Lower Half of the Microscope 12 1 Focus Knob 12 4 DIC Analyzer slider out position 12 3 Lens Drop Raise 12 6 DIC Anal...

Page 41: ...rference Contrast DIC This enables high contrast high resolution images by transmitted light from samples that are only stained with fluorescent labels Using DIC greatly reduces photobleaching while l...

Page 42: ...ton prism into place 7 3 Setting Koehler Illumination Maximal resolution and field uniformity with brightfield and DIC is only achieved if the condenser is in focus to the same plane as the objective...

Page 43: ...ht intensity is restored 9 Check the focus of the sample it may change if the condenser was initially out of focus 10 If the focus was changed re check the condenser by simply closing the field stop a...

Page 44: ...om to a value greater than Zoom 1 concentrates the laser power of into a much smaller volume The result is that the photodynamic effects are achieved with much lower laser level and shorter exposure d...

Page 45: ...umber of scans or a defined time period ms f Activation in Series allows photoactivation operations in a sequence i PreActivation record a specified number of images before activation ii Activation Ti...

Page 46: ...cence to confocal imaging Rotating the mirrors filter cubes to allow confocal scanning exposes the entire field of view to a brief flash of UV from the DAPI cube Close the manual shutter below the obj...

Page 47: ...em redo work space color Figure 17 2 4 Be aware that menus and toolbar items are nearly impossible to see in Darkroom color mode 17 1 Darkroom Color 17 2 Redo Work Space Figure 17 The toolbar item for...

Page 48: ...s e g Cargille Type B for the least spread of oil 8 Do not over oil immersion objectives use the minimum amount of oil necessary a Scrunchies have been wrapped around the oil immersion objectives to a...

Page 49: ...hrough dichroics measured at the lens turret 100 output Dichroic Mirror 405 nm 458 nm 488 nm 515 nm 561 nm 633 nm BS20 80 161 W 9 4 W 23 5 W 50 W 45 W 98 W BM405 488 414 W 1 W 102 W 16 W 1 2 W 20 4 W...

Page 50: ...ncreased Has the detector Offset been increased This should be 5 7 Are you adjusting to the correct PMT and laser Does the Confocal Aperture need to be opened up Do you need to use special imaging met...

Reviews: