86
Appendix
EVIS EXERA II TJF TYPE Q180V OPERATION MANUAL
Guidance and manufacturer’s declaration —
Electromagnetic immunity
This model is intended for use in the electromagnetic environment specified
below. The customer or the user of this model should assure that it is used
in such an environment.
Portable and mobile RF communications equipment should be used no
closer to any part of this model, including cables, than the recommended
separation distance calculated from the equation applicable to the frequency
of the transmitter.
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
• Electromagnetic interference may occur in the vicinity of
high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment — Guidance
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
(150 kHz – 80 MHz)
3 V (V
1
)
Recommended separation distance
Radiated RF
IEC 61000-4-3
3 V/m
(80 MHz – 2.5 GHz)
3 V/m (E
1
)
80 MHz – 800 MHz
800 MHz – 2.5 GHz
Definition:
Where “P” is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer and “d” is the recommended separation distance in meters (m).
d
3,5
V
1
-------
P
=
d
3,5
E
1
-------
P
=
d
7
E
1
------
P
=
Summary of Contents for EVIS EXERA II